Table 2.

Pseudo replica coefficients of variation (CVs) and within-session repeatability of MRF T1 and T2 measurements in 4 healthy volunteers. For each tissue type, T1 and T2 was calculated from 3D regions-of-interest. The mean ± standard deviation (SD) across subjects is reported for the repeated measures CV and the Bland-Altman bias is calculated from the group.

Pseudo-replica CV Repeated measures CV (mean ± SD) Bland Altman Bias ± 1.96 SD
White Matter T1 0.008% 0.76 ± 0.35% 1.5 ± 12 ms
T2 0.014% 3.8 ± 5.0% −1.9 ± 6.1 ms
Gray Matter Tl 0.007% 0.7 ± 0.5% 5.0 ± 19.1 ms
T2 0.009% 3.8 ± 5.2% −2.8 ± 10.9 ms
Cerebrospinal fluid T1 0.011% 2.5 ± 1.7% 6.0 ± 236.1 ms
T2 0.028% 3.5 ± 0.6% 1.4 ± 57.9 ms