Inter- and intrarun, interlot, and intersite reproducibilities of viral load assays

Reproducibility Parametera Result at the indicated siteb for the following assay:
usCA MONITOR 1.5
stCA MONITOR 1.5
NucliSens
bDNA 3.0
Site A Site B Combined Site A Site B Combined Site A Site C Combined Site B Site C Combined
Inter- and intrarun SEM 0.14 0.12 0.13 0.21 0.17 0.19 0.22 0.14 0.18 0.11 0.06 0.09
% CV 28.8 25.0 26.9 0.22 37.0 38.2 40.9 22.2 31.2 22.5 13.1 18.2
n 23 24 47 0.23 19 41 28 30 58 28 24 52
Interlot SEM 0.12 0.13 0.13 0.14 0.12 0.13 0.19 0.12 0.16 0.13 0.09 0.12
% CV 23.0 24.1 23.6 21.2 21.4 21.3 30.1 22.5 26.1 25.0 17.6 21.9
n 24 25 49 24 20 44 27 26 53 33 24 57
Intersite SEM 0.14 0.20 0.22 0.12
% CV 23.8 30.5 29.6 21.3
n 51 44 54 55
a

SEM was calculated from log10 RNA copies per milliliter. Percent CV is the mean value calculated from RNA copies per milliliter. n is number of samples included in the analysis. 

b

Sites A, B, and C are the three hospital test sites.