. 2000 Nov;38(11):4171–4179. doi: 10.1128/jcm.38.11.4171-4179.2000

TABLE 2.

Limit of detection for the COBAS AMPLICOR and AMPLICOR HCV MONITOR version 2.0 testsa

Actual input (IU/ml) % Positive (no. positive/total no. of tests)
COBAS AMPLICOR format AMPLICOR format
1,100 100 (48/48) 100 (48/48)
900 95 (91/96) 99 (94/95)b
700 99 (95/96) 100 (96/96)
600 95 (74/78) 96 (75/78)
500 90 (92/102) 98 (106/108)
400 97 (29/30) 89 (32/36)
300 57 (17/30) 75 (27/36)
a

Limit of detection was determined with dilutions of the secondary standard that was calibrated against the WHO international standard for HCV (96/790) as described in Results. 

b

One reaction was excluded from the calculation because it gave a negative IC result.