Table 1.
NCP M100* | NCP M101 | Pulse M102™ | Pulse Duo™ M102R | Demipulse® M103 | Demipulse® Duo M104 | AspireHC® M105 | AspireSR® M106 | SenTiva® M1000 | SenTiva Duo™ M1000-D† | |
---|---|---|---|---|---|---|---|---|---|---|
Lead receptacle | Dual-pin | Dual-pin | Single-pin | Dual-pin | Single-pin | Dual-pin | Single-pin | Single-pin | Single-pin | Dual-pin |
FDA release year | 1997 | 2000 | 2002 | 2003 | 2007 | 2007 | 2011 | 2015 | 2017 | ——– |
EU certification mark year | 1994 | 1999 | 2003 | 2003 | 2005 | 2005 | 2011 | 2014 | 2017 | 2020 |
HS-Ti case dimension (mm) | 55 ×55 ×13.2 | 54 ×54 ×10.3 | 52 ×52 ×6.9 | 52 ×58.4 ×6.9 | 45 ×32 ×6.9 | 45 ×39 ×6.6 | 52 ×52 ×6.9 | 52 ×52 ×6.9 | 45 ×32 ×6.9 | 45 ×39 ×6.9 |
Volume (cc) | 31 | 26 | 14 | 16 | 8 | 10 | 14 | 14 | 8 | 10 |
Weight (g) | 55 | 38 | 25 | 27 | 16 | 17 | 25 | 25 | 16 | 17 |
Battery model | 9086 | 9086 | 2075 | 2075 | 2183 | 2183 | 2075 | 2075 | 2183 | 2183 |
Rated capacity (amp-hours) | 2.3 | 2.3 | 1.7 | 1.7 | 1.0 | 1.0 | 1.7 | 1.7 | 1.0 | 1.0 |
X-ray tag code | None | None | CYBX or CYBX-J-XX | CYBX or CYBX-J-XX | CYB A & VNS A | CYB A & VNS A | CYBX | CYBX | LIVN & VNS | LIVN & VNS |
Time needed for magnet inhibition (s) | ≥ 65 | ≥ 65 | ≥ 65 | ≥ 65 | ≥ 65 | ≥ 65 | ≥ 65 | ≥ 5 | ≥ 10 | ≥ 10 |
Feature or specification (left) with respect to VNS Therapy generator model (top). For x-ray tag code CYBX-J-XX, the “XX” signifies the year (e.g., 20 for 2020). Time needed for magnet inhibition is the time the therapy magnet needs to be held over the generator to stop stimulation.
Applies to M100 generators with serial numbers ≥10,000.
M1000-D generator is only licensed in Europe as of this review.
HC, high capacity; HS-Ti, hermetically sealed titanium; NCP, NeuroCybernetic Prosthesis; M, model; SR, sense and respond.