Table 1.

Feature and specification table for VNS Therapy generator models.

NCP M100* NCP M101 Pulse M102™ Pulse Duo™ M102R Demipulse® M103 Demipulse® Duo M104 AspireHC® M105 AspireSR® M106 SenTiva® M1000 SenTiva Duo™ M1000-D
Lead receptacle Dual-pin Dual-pin Single-pin Dual-pin Single-pin Dual-pin Single-pin Single-pin Single-pin Dual-pin
FDA release year 1997 2000 2002 2003 2007 2007 2011 2015 2017 ——–
EU certification mark year 1994 1999 2003 2003 2005 2005 2011 2014 2017 2020
HS-Ti case dimension (mm) 55 ×55 ×13.2 54 ×54 ×10.3 52 ×52 ×6.9 52 ×58.4 ×6.9 45 ×32 ×6.9 45 ×39 ×6.6 52 ×52 ×6.9 52 ×52 ×6.9 45 ×32 ×6.9 45 ×39 ×6.9
Volume (cc) 31 26 14 16 8 10 14 14 8 10
Weight (g) 55 38 25 27 16 17 25 25 16 17
Battery model 9086 9086 2075 2075 2183 2183 2075 2075 2183 2183
Rated capacity (amp-hours) 2.3 2.3 1.7 1.7 1.0 1.0 1.7 1.7 1.0 1.0
X-ray tag code None None CYBX or CYBX-J-XX CYBX or CYBX-J-XX CYB A & VNS A CYB A & VNS A CYBX CYBX LIVN & VNS LIVN & VNS
Time needed for magnet inhibition (s) ≥ 65 ≥ 65 ≥ 65 ≥ 65 ≥ 65 ≥ 65 ≥ 65 ≥ 5 ≥ 10 ≥ 10

Feature or specification (left) with respect to VNS Therapy generator model (top). For x-ray tag code CYBX-J-XX, the “XX” signifies the year (e.g., 20 for 2020). Time needed for magnet inhibition is the time the therapy magnet needs to be held over the generator to stop stimulation.

*

Applies to M100 generators with serial numbers ≥10,000.

M1000-D generator is only licensed in Europe as of this review.

HC, high capacity; HS-Ti, hermetically sealed titanium; NCP, NeuroCybernetic Prosthesis; M, model; SR, sense and respond.