Skip to main content
. 2022 Jan 14;116:358–364. doi: 10.1016/j.ijid.2022.01.019

Table 2.

Self-RADT performance

Overall TP FN FP TN Sensitivity (95% CI) Specificity (95% CI)
Hangzhou 401 41 25 0 335 62.1 (50.1–72.9) 100 (98.9–100)
DNA Diagnostic 423 23 12 0 388 65.7 (49.2–79.2) 100 (99.0–100)
Abbott 388 0 0 1 387 Not estimable 100 (95.6–100)

Six RADTs (n = 1 DNA Diagnostic, n = 2 Hangzhou, and n = 3 Abbott) were not included in the sensitivity and specificity calculations. RADT results that differed from the PCR test results were considered FP or FN; concurrences between the 2 tests were considered TP or TN.