Table 1.

Accuracy and interday precision estimated in standard medium and biological samples

  In IDO1 activity mixture
In cell lysatea
ʟ-Kyn concentration added (µM)
0.043 2.15 8.60 0.043b 2.15b 8.60b
Day 1
 Mean ± SD (µM) 0.039 ± 0.002 2.193 ± 0.017 8.822 ± 0.070 0.048 ± 0.005 2.428 ± 0.008 9.524 ± 0.029
 RSD (%) 6.16 0.76 0.79 5.31 0.32 0.31
 Accuracy (%) 90.60 101.98 103.74 112.67 112.95 110.74
Day 2
 Mean ± SD (µM) 0.040 ± 0.003 2.034 ± 0.007 8.586 ± 0.026 0.043 ± 0.010 1.982 ± 0.013 8.204 ± 0.023
 RSD (%) 8.07 0.34 0.30 2.03 0.53 0.33
 Accuracy (%) 93.27 94.59 99.84 111.03 89.92 95.40
Day 3
 Mean ± SD (µM) 0.039 ± 0.002 1.996 ± 0.007 8.256 ± 0.021 0.040 ± 0.006 2.102 ± 0.181 8.936 ± 0.353
 RSD (%) 5.03 0.37 0.26 6.53 8.41 3.93
 Accuracy (%) 90.39 92.84 96.00 91.94 97.77 103.91

aCell lysate from MDA-MB-231 cancer cells; bcalculated as follows: ʟ-Kyn concentration in spiked sample (µM) – ʟ-Kyn concentration in blank sample (µM).

Standard deviation (SD) values were calculated from 6 measurements.