Clinical performance of Roche Elecsys SARS-CoV-2 Antigen fully automated electrochemiluminescence immunoassay (ECLIA) stratified according to cycle threshold (Ct) values.

Ct values AUC Sensitivity Specificity



All samples 0.83 (95%CI, 0.77–0.89; p < 0.001 0.43 (95%CI, 0.333–0.525) 1.00 (95%CI, 0.93–1.00)



<30 0.99 (95%CI, 0.98–1.00; p < 0.0001) 0.87 (95%CI, 0.75–0.945) 0.99 (95%CI, 0.95–1.00)



<25 1.00 (95%CI, 0.99–1.00; p < 0.001) 1.00 (95%CI, 0.89–1.00) 0.89 (95%CI, 0.82–0.94)




AUC, area under the curve; Ct, cycle threshold.