Skip to main content
. 2022 Jan 3;11:784610. doi: 10.3389/fcimb.2021.784610

Table 4.

Results of the validation for the discordant results detected by Cepheid GeneXpert® (Xpert) and Roche cobas® 4800 CT/NG tests for urogenital Chlamydia trachomatis (CT) and Neisseria gonorrhoeae (NG) .

Sequencing results CT Sequencing results NG
Xpert positive and Roche negative Xpert negative and Roche positive Xpert positive and Roche negative Xpert negative and Roche positive
CT-urine Positive 1 0 NG-urine Positive 1 1
Negative 8 2 Negative 0 4
CT-swab Positive 11 1 NG-swab Positive 3 0
Positive 12 8 Negative 3 0