TABLE 5.
Inter‐rater reliability between wound nurse and WoundAide device 1 for the corresponding image
Measurements, mean ± SD | Inter‐rater reliability | P value | |||||||
---|---|---|---|---|---|---|---|---|---|
Length | Width | Area | Length | Width | Area | Length | Width | Area | |
Device 1 | 3.43 ± 2.49 | 2.49 ± 1.70 | 9.03 ± 12.47 | 0.882 (0.829, 0.916) | 0.893 (0.860, 0.918) | 0.936 (0.899, 0.957) | <.001 | <.001 | <.001 |
Wound nurse | 3.86 ± 2.82 | 2.67 ± 1.94 | 10.95 ± 16.36 |
Note: Length and breadth are expressed in cm. Area was expressed in cm2.