Table 3.
Diagnostic capability of different anterior segment optical coherence tomography (AS-OCT) and Pentacam parameters. AUC: area under the ROC curve (receiver operating curves) and 95% confidence interval. Cut-off values, sensitivity (%), and specificity (%). AS-OCT parameters: perpendicular angle opening distance from Schwalbe line (SL) to iris (AOD-SL-Perp temporal and nasal), vertical angle opening distance from Schwalbe line (SL) to iris (AOD-SL-Vert temporal and nasal), and iridotrabecular angle (ITA temporal and nasal). Pentacam anterior segment parameters: anterior chamber depth (ACD), anterior chamber volume (ACV), and anterior chamber angle (ACA)
Diagnostic capability AS-OCT: | AUCs | 95% CI AUC | Cut-off value | Sensitivity (%) | Specificity (%) |
---|---|---|---|---|---|
AOD-SL-Perp temporal (microns) | 0.898 | 0.802 to 0.957 | 230 | 90.91 | 73.33 |
AOD-SL-Perp nasal (microns) | 0.834 | 0.724 to 0.913 | 194 | 90.11 | 53.33 |
AOD-SL-Vert temporal (microns) | 0.874 | 0.773 to 0.941 | 273 | 76.36 | 93.33 |
AOD-SL-Vert nasal (microns) | 0.876 | 0.774 to 0.944 | 205 | 92.45 | 73.33 |
ITA temporal (degrees) | 0.886 | 0.783 to 0.951 | 26 | 60.78 | 100 |
ITA nasal (degrees) | 0.913 | 0.818 to 0.968 | 23 | 78.43 | 100 |
Diagnostic capability Pentacam | AUCs | 95% CI AUC | Cut-off value | Sensitivity (%) | Specificity (%) |
ACD (mm) | 0.764 | 0.648 to 0.857 | 2.16 | 92.45 | 64.71 |
ACV (mm3) | 0.754 | 0.636 to 0.849 | 97 | 90.57 | 64.71 |
ACA (degrees) | 0.822 | 0.712 to 0.903 | 26.5 | 84.91 | 70.59 |