Comparison of tests for the identification of S. aureusa

Test and centerb Sensitivity (%)
Specificityc
MSSA MRSA All isolates
Slidex Staph Plus 98.9 97.1 98.2 98.9
 Zurich 98.0 95.9 97.3 99.3
 Lyon 98.7 97.3 98.0 98.7
 Tilburg 100 98.0 99.3 98.7
Staphaurex Plus 98.2 98.2 98.2 96.2
 Zurich 98.0 100 98.7 95.1
 Lyon 96.1 97.3 96.7 96.0
 Tilburg 100 98.0 99.3 97.4
Pastorex Staph-Plus 98.6 98.8 98.7 95.7
 Zurich 98.0 100 98.7 95.8
 Lyon 98.7 97.3 98.0 95.3
 Tilburg 99.0 100 99.3 96.0
Tube coagulase 99.3 100 99.6 99.5
 Zurich 99.0 100 99.3 99.3
 Lyon 100 100 100 100
 Tilburg 99.0 100 99.3 99.3
a

Values were determined with 892 staphylococcal isolates, composed of 449 S. aureus (171 methicillin resistant) and 443 CoNS isolates. 

b

Zurich, Department of Clinical Microbiology, University of Zurich, Zurich, Switzerland; Lyon, Laboratory of Microbiology Hospital Edouard Herriot, Lyon, France; Tilburg, Department of Clinical Microbiology, St. Elisabeth Hospital, Tilburg, The Netherlands. 

c

The specificity of Slidex Staph Plus is statistically significantly higher than those of Staphaurex Plus and Pastorex Staph-Plus (P < 0.05).