Table 3.

Characteristics of chip-based electrochemical O2 sensors.

EC-Based O2 Sensor LOD Advantages References
PDMS-container structure, and the glass substrate 105 cells/mL Short response time (6.9 s) [83]
Low-temperature co-fired ceramic (LTCC) in an improved Clark-type DO sensor Up to 8.1 mg/L easy fabrication, flexible configuration, short response time (10.9 s), real-time detection [84]
pHEMA hydrogel layer with electrolyte and PDMS as gas-permeable membrane 0.121 μA cm−2 μM−1 zero analyte consumption, 1-point calibration, long-term stability [85]
PPy as the internal contact layer between polymeric sensitive membrane and gold 0.11 ± 0.02 mg L−1 Low cost, good performance and long-term potential stability [86]
Multi-sensor glass-chip with a PDMS imprinted microfluidic channel grid 100 pA per each 1% O2 Transparent for microscopic observation, cheap, high sensitivity [87]
Biocompatible glass chip fabricated using a hybrid thin film and laminate technologies 0.735 μA μM−1 cm−2 Low O2 consumption on the electrode, long-term stability [88]
Biocompatible PDMS biochip with Au/Nafion electrodes 50 mmol L−1 real-time and continuous O2 monitoring in dynamic flow conditions [89]
Kapton tape with embedded spirally rolled Microchannels 12.89 nA mmHg −1 O2 and temperature sensors, embedded spirally rolled microchannels [90]
ElecCell technological platform using PVD 6 pA/s low-cost, easy to use and reproducible portable chip [91]
ultra-microelectrode array (UMEA) 0.49 nAs−0.5/mg/L Ultra-short response time (<5 ms), 10 times lower O2 consumption [92]
Multi-planar SPE sensor coupled with cultivation cell wells 3 mg/L Continuous long-term O2 measurement, sensor reutilization [93]
Inkjet printing (IJP) DO sensors on the delicate porous substrate 28 ± 1 nA L mg−1 low O2 consumption on electrodes, short response time (60 s) [94]
Electrochemical microsensors combined with spheroid technology NM fast, precise, and continuous long-term measurement of metabolic directly in the microwell [95]
Spheroid on chip NM Real-time monitoring of metabolic activity and automated assays for toxicity evaluation [96]

NM: not mentioned.