Table 1.

Comparison of Allplex RP 1, 2, 3 and BioFire FilmArray RP assays for the detection of respiratory viruses in NPS specimens.

Viruses Allplex RP 1, 2, and 3 BioFire FilmArray RP
Positive Negative Positive Negative
AdV 31 (3) 150 34 (5) 147
229E 0 181 0 181
NL63 0 181 0 181
OC43 0 181 0 181
HKU1 NT NT 0 181
FluA 2 179 2 179
FluB 0 181 0 181
HBoV-1/2/3/4 10 171 NT NT
HEV/HRV a 64 (25) 117 39 142
HEV 24 (12) 157 NT NT
HRV 55 (26) 126 NT NT
MPV 0 181 0 181
PIV-1 6 (3) 175 4 (1) 177
PIV-2 5 (3) 176 2 179
PIV-3 5 (1) 176 6 (2) 175
PIV-4 6 (4) 175 2 179
RSV-A/RSV-B b 4 (1)/4 177/177 8 (1) 173
RSV-A 4 (1) 177 NT NT
RSV-B 4 177 NT NT

a: BioFire FilmArray RP assay does not differentiate between human enterovirus (HEV) and human rhinovirus (HRV) and reports them as a single analyte, HEV/HRV. b: BioFire FilmArray RP assay does not differentiate between RSV-A and RSV-B and reports them as a single analyte RSV-A/RSV-B. Numbers in parentheses indicate the analytes detected only by the one RP assay. NT: not tested. Abbreviations: AdV, adenovirus; 229E, NL63, and OC43, coronavirus; FluA, influenza A virus; FluB, influenza B virus; HBoV, human bocavirus 1/2/3/4; HEV, human enterovirus; MPV, human metapneumovirus; HRV, human rhinovirus; PIV-1, PIV-2, PIV-3, and PIV-4, parainfluenza viruses 1, 2, 3, and 4; RSV-A and RSV-B, respiratory syncytial virus A and B.