Table 1.
SNR, whole-lung mean, and whole-lung standard deviation for ventilation, RBC, and Barrier images of healthy subjects. Values are expressed as the cohort mean ± standard deviation. Given similar xenon dose equivalents were given for standard and fast acquisitions (152 and 159 mL), the differences in SNR due to xenon polarization are expected to be minimal. Abbreviations: Vent – Ventilation Image, SNR – Signal to Noise Ratio, RBC – Red Blood Cell
Scan | Vent SNR | Total Dissolved SNR | Barrier SNR | RBC SNR | Vent Mean | Barrier Mean | RBC Mean | Vent Standard Deviation | Barrier Standard Deviation | RBC Standard Deviation |
---|---|---|---|---|---|---|---|---|---|---|
Standard | 15±4 | 23±8 | 15±5 | 8±3 | 0.43±0.07 | 0.73±0.10 | 0.45±0.07 | 0.20±0.02 | 0.22±0.11 | 0.25±0.13 |
Fast | 18±5 | 15±6 | 9±4 | 5±2 | 0.41±0.07 | 0.72±0.13 | 0.45±0.08 | 0.22±0.03 | 0.24±0.08 | 0.28±0.09 |
Fast1500 | 20±5 | 19±7 | 12±4 | 6±3 | 0.41±0.08 | 0.73±0.13 | 0.44±0.09 | 0.22±0.03 | 0.21±0.08 | 0.24±0.07 |
Fast1000 | 21±4 | 23±8 | 15±5 | 8±3 | 0.41±0.07 | 0.73±0.14 | 0.44±0.09 | 0.22±0.03 | 0.18±0.04 | 0.20±0.05 |