Technical aspects of the POLARx compared to the AFA system.

Characteristics POLARx AFA
Sheath diameter (F) 12.7 12
Sheath outer diameter (F) 15.9 15
Radiopaque marker proximal to the tip (mm) 2.5 5
Balloon size (mm) 28 28
Balloon shaft diameter (F) 10.5
Balloon tip length (mm) 5 or 12 8
N2O injection 8-hole coil 8-hole coil
N2O fluid flow during freeze (sccm) 7800 7200
Pressure during freeze (psi) <525 constant 530–600
Location of injection coil from pole of balloon (mm) 2.5 3.5
Location of TC from coil (mm) 18 15
Location of gas outflow proximal of TC (mm) 5 10
Phrenic nerve palsy control DMS (integrated/quantitative) CMAP (not integrated/not quantitative)
Console register procedural data yes no
Console operation autonomically yes no

CMAP, compound motor action; DMS, diaphragm movement sensor; TC, thermocouple.