Skip to main content
. 2021 Dec 31;8(1):4. doi: 10.3390/jimaging8010004

Table 2.

Comparison of values for several key design parameters, as used in leading detectors and a simulation-optimised system. Charge-sharing correction algorithm (CSCA) parameters of neighbourhood size (NS) and neighbourhood location (NL) are looked at here and defined in other publications [76].

Optimised Detector MARS Detector Phillips Healthcare Detector
Pixel pitch (µm) 200–250 110 500
Sensor thickness (mm) ~1.5 2 2
CSCA NS 3 × 3 or Hybrid Hybrid 1 × 1 or 2 × 2 (chequered)
CSCA NL Dynamic Dynamic Static