Overview of precursor ion, product ions and limit of detection for not validated microcystin congeners. Additionally, the table also includes the detection frequency of the congeners in the analyzed samples at different total microcystin concentrations.

Toxins MC-HtyR dm MC-LR/D-asp MC-LR D-asp-Dhb MC-RR/dm MC-RR MC-HilR
Precursor ion 1059.5 981.14 512.7 505.3
Product ions 106.9;
135.27
106.8;
135.07
103.2;
135.13
126.99;
134.92
Limit of Detection (µg L−1) 0.1 0.1 0.1 0.1
All samples 13.92% 53.16% 77.22% 34.18%
Samples < 1 µg L−1 total microcystin 0.00% 9.38% 50.00% 0.00%
Samples > 1 µg L−1 total microcystin 23.40% 82.98% 95.74% 57.45%
Samples < 24 µg L−1 total microcystin 25.00% 100.00% 100.00% 100.00%