Skip to main content
. 2021 Dec 22;13(12):e20627. doi: 10.7759/cureus.20627

Table 6. Determination of test accuracy using Sansure, STAT-NATⒷ, and Roche RT-qPCR kits.

CI: confidence interval; Sn: sensitivity; Sp: specificity; LR+: likelihood ratio for a positive test; LR-: likelihood ratio for a negative test; RT-qPCR: quantitative reverse transcription-polymerase chain reaction

  Diagnostic accuracy with 95% CI
RT-qPCR kit Sn (%) Sp (%) Overall acuracy (%) LR+ LR-
Sansure 62.24 (51.88-71.84) 96.74 (90.77-99.32) 78.95 (72.46-84.51) 19.09 (6.21-58.72) 0.39 (0.30-0.50)
STAT-NAT 83.56 (73.05-91.21) 97.80 (92.29-99.73) 91.46 (86.09-95.25) 38.02 (9.62-150.28) 0.17 (0.10-0.28)
Roche 96.83 (89.00-99.61) 89.00 (81.17-94.38) 93.17 (88.10-96.54) 8.80 (5.03-15.40) 0.04 (0.01-0.14)