Table 6. Determination of test accuracy using Sansure, STAT-NATⒷ, and Roche RT-qPCR kits.
CI: confidence interval; Sn: sensitivity; Sp: specificity; LR+: likelihood ratio for a positive test; LR-: likelihood ratio for a negative test; RT-qPCR: quantitative reverse transcription-polymerase chain reaction
Diagnostic accuracy with 95% CI | |||||
RT-qPCR kit | Sn (%) | Sp (%) | Overall acuracy (%) | LR+ | LR- |
Sansure | 62.24 (51.88-71.84) | 96.74 (90.77-99.32) | 78.95 (72.46-84.51) | 19.09 (6.21-58.72) | 0.39 (0.30-0.50) |
STAT-NATⒷ | 83.56 (73.05-91.21) | 97.80 (92.29-99.73) | 91.46 (86.09-95.25) | 38.02 (9.62-150.28) | 0.17 (0.10-0.28) |
Roche | 96.83 (89.00-99.61) | 89.00 (81.17-94.38) | 93.17 (88.10-96.54) | 8.80 (5.03-15.40) | 0.04 (0.01-0.14) |