Skip to main content
. 2022 Jan 10;11:812643. doi: 10.3389/fonc.2021.812643

Table 4.

Global gamma evaluation, with a dose cut-off at 15%, comparing sCT and CT dose distributions, averaged over all patients.

Gamma criteria Gamma pass rate ± 1 SD [range] (%)
3%, 3 mm 100.0 ± 0.1 [99.7–100.0]
2%, 2 mm 99.8 ± 0.2 [99.4–100.0]
1%, 1 mm 99.1 ± 0.6 [97.6–99.8]