Table 2: Recommended Programming of Closed-loop Stimulation in Cardioinhibitory Reflex Syncope.

Programmable parameter Value
Mode DDD-CLS
Basic rate day/night 50 BPM/OFF
Maximum CLS rate 120 BPM
CLS response Medium
CLS resting rate control OFF
Atrioventricular delay Low; 180–140 ms
Atrioventricular hysteresis mode IRSplus
Atrial overdrive pacing OFF
PMT protection ON
IEGM recordings
High atrial rate AT
High ventricular rate ON
Patient triggering OFF
Rates for statistics
High atrial rate limit 200 BPM
High ventricular rate limit 180 BPM
High ventricular rate counter 8 events

AT = atrial tachyarrhythmia; CLS = closed-loop stimulation; DDD = dual-chamber sensing and pacing; IEGM = intracardiac electrogram; IRS = intrinsic rhythm support (a specific function favouring intrinsic atrioventricular contraction); PMT = pacemaker-mediated tachycardia. Source: Brignole et al.[6]