Table 2: Recommended Programming of Closed-loop Stimulation in Cardioinhibitory Reflex Syncope.
| Programmable parameter | Value |
|---|---|
| Mode | DDD-CLS |
| Basic rate day/night | 50 BPM/OFF |
| Maximum CLS rate | 120 BPM |
| CLS response | Medium |
| CLS resting rate control | OFF |
| Atrioventricular delay | Low; 180–140 ms |
| Atrioventricular hysteresis mode | IRSplus |
| Atrial overdrive pacing | OFF |
| PMT protection | ON |
| IEGM recordings | |
| High atrial rate | AT |
| High ventricular rate | ON |
| Patient triggering | OFF |
| Rates for statistics | |
| High atrial rate limit | 200 BPM |
| High ventricular rate limit | 180 BPM |
| High ventricular rate counter | 8 events |
AT = atrial tachyarrhythmia; CLS = closed-loop stimulation; DDD = dual-chamber sensing and pacing; IEGM = intracardiac electrogram; IRS = intrinsic rhythm support (a specific function favouring intrinsic atrioventricular contraction); PMT = pacemaker-mediated tachycardia. Source: Brignole et al.[6]