Skip to main content
. 2022 Jan 17;2022:5952436. doi: 10.1155/2022/5952436

Table 2.

Intrabatch (n = 5) and interbatch (n = 15) accuracy and precision of trans-resveratrol, R3S, and R3G in mice blood at four QC levels.

Analytes Nominal concentration (ng/mL) Intrabatch (n = 5) Interbatch (n = 15)
Measured concentration (ng/mL) Precision (%) Accuracy (%) Measured concentration (ng/mL) Precision (%) Accuracy (%)
Trans-resveratrol 2.24 2.30 ± 0.25 10.91 103.23 2.31 ± 0.20 8.81 103.71
5.59 5.93 ± 0.56 9.45 106 5.52 ± 0.50 9.11 98.65
22.35 25.24 ± 0.47 1.87 112.6 23.69 ± 2.13 9.00 105.69
111.75 108 ± 4.52 4.18 96.28 107.34 ± 9.02 8.40 95.79

R3G 1.00 1.07 ± 0.15 14.00 106.9 1.03 ± 0.13 12.22 103.44
5.00 5.25 ± 0.30 6.00 104.9 5.33 ± 0.32 6.07 106.46
20.00 18.1 ± 0.91 5.00 90.48 18.31 ± 1.25 6.86 91.57
100.00 90.7 ± 4.67 5.00 90.7 96.21 ± 7.74 8.04 96.21

R3S 1.00 1.02 ± 0.08 8.11 101.50 0.96 ± 0.09 9.38 96.05
5.00 5.19 ± 0.22 4.29 103.87 5.34 ± 0.37 6.97 106.83
20.00 18.04 ± 0.61 3.00 90.30 18.67 ± 1.65 8.84 93.33
100.00 103.07 ± 9.13 8.86 103.07 100.73 ± 10.96 10.88 100.73

Data of measured concentration represented as mean ± SD. R3G, trans-resveratrol-3-o-β-glucuronide; R3S, trans-resveratrol-3-sulfate salt.