Table 7.

Comparison of mechanical circulatory support system

IABP IMPELLA VA-ECMO
Cardiac flow 0.3–0.5 L/min 1–5 L/min (Impella 2.5, Impella CP, Impella 5) 3–7 L/min
Mechanism Aorta LV → Ao RA → Ao
Maximum implant days Weeks 7 days Weeks
Sheath size 7–8 Fr

13–14 Fr

Impella 5.0—21 Fr

14–16 Fr arterial

18–21 Fr venous

Femoral artery size  > 4 mm

Impella 2.5 and CP: 5–5.5 mm

Impella 5: 8 mm

8 mm
Cardiac synchrony or stable rhythm Yes No No
Afterload ↑↑↑
Mean arterial pressure ↑↑ ↑↑
LVEDP ↓↓
PCWP ↓↓
LV preload ↓↓
Coronary perfusion
Myocardial oxygen demand ↓↓

Modified from [167]

Ao aorta, IABP intra-aortic balloon pump, LA left atrium, LV left ventricle, LVEDP left ventricular end diastolic pressure, RA right atrium, PCWP pulmonary capillary wedge pressure, VA-ECMO  venoarterial extracorporeal membrane oxygenation