Integrated deep learning framework for accelerated optical coherence tomography angiography

Gyuwon Kim 1,#, Jongbeom Kim 1,2,#, Woo June Choi 3,, Chulhong Kim 1,2,4,, Seungchul Lee 1,4,
PMCID: PMC8789830  PMID: 35079046

Abstract

Label-free optical coherence tomography angiography (OCTA) has become a premium imaging tool in clinics to obtain structural and functional information of microvasculatures. One primary technical drawback for OCTA, however, is its imaging speed. The current protocols require high sampling density and multiple acquisitions of cross-sectional B-scans to form one image frame, resulting in low acquisition speed. Recently, deep learning (DL)-based methods have gained attention in accelerating the OCTA acquisition process. They achieve faster acquisition using two independent reconstructing approaches: high-quality angiograms from a few repeated B-scans and high-resolution angiograms from undersampled data. While these approaches have shown promising results, they provide limited solutions that only partially account for the OCTA scanning mechanism. Herein, we propose an integrated DL method to simultaneously tackle both factors and further enhance the reconstruction performance in speed and quality. We designed an end-to-end deep neural network (DNN) framework with a two-staged adversarial training scheme to reconstruct fully-sampled, high-quality (8 repeated B-scans) angiograms from their corresponding undersampled, low-quality (2 repeated B-scans) counterparts by successively enhancing the pixel resolution and the image quality. Using an in-vivo mouse brain vasculature dataset, we evaluate our proposed framework through quantitative and qualitative assessments and demonstrate that our method can achieve superior reconstruction performance compared to the conventional means. Our DL-based framework can accelerate the OCTA imaging speed from 16 to 256× while preserving the image quality, thus enabling a convenient software-only solution to enhance preclinical and clinical studies.

Subject terms: Imaging and sensing, Biophotonics

Introduction

Optical coherence tomography (OCT) is an essential biomedical imaging technique based on low-coherence interferometry to image biological tissues. It enables non-invasive, depth-resolved imaging by demodulating backscattered light interference1,2. The ability to resolve axial structure superiorly has led to various clinical applications, such as neurology, ophthalmology, dermatology, gastroenterology, and cardiology35. Furthermore, OCT has also been extended to functional angiography, namely, optical coherence tomography angiography (OCTA), to study vascular diseases. OCTA extracts the variations in OCT signals caused by the red blood cell (RBC) motion to contrast blood flow within vessels against the static tissue. This functional angiography is accomplished using a sequence of repeated B-scans taken from a single cross-sectional location6. The agent-free OCTA modality, owing to its micro-level spatial resolution and non-invasive nature, has enabled anatomical and functional imaging in various clinical applications710. However, one limitation of OCTA is acquisition speed, bound by the scanning principle employed, i.e., point-by-point raster scanning and the requirement of sequential B-scans. Since high-speed imaging is very desirable for functional imaging, a branch of study has emerged focusing on developing OCT systems with ultrahigh imaging speed, e.g., high-speed optical sources1113 and detection devices1416. However, enahncing imaging speed using hardware adjustments is cost-prohibitive and may pose limitations for practical clinical system implementations.

Recently, several studies have been conducted to enhance the OCTA imaging speed via advances of emerging deep learning (DL)-based methods. Such studies aim to provide software-only solutions that require no modifications to the hardware settings. A study in17 adopted convolutional neural network (CNN) architectures to accelerate the B-scan repetition procedure by reconstructing high-quality cross-sectional angiograms (acquired from 48 consecutive B-scans) from a limited number of B-scans (n 4). For the resulting maximum intensity projection (MIP) enface angiograms, the study reported structural similarity index measure (SSIM) and peak signal-to-noise ratio (PSNR) values of 0.63 and 20.82, respectively. Another work in Ref.18 conducted pixel-wise super-resolution in the enface plane using deep learning to reduce the sampling density while preserving the pixel resolution. Results have outperformed baseline filter-based methods and showed improved signal-to-noise ratio (SNR) and vascular connectivity.

Although DL-based methods have demonstrated the effectiveness in accelerating the OCTA procedure, the following aspects require further investigation, which has motivated the current study. First, all factors related to the imaging speed (i.e., the B-scan repetition number and the sampling density) should be considered simultaneously in a single framework. The previous works17,18 addressed only one of the mentioned factors. Secondly, a DL-framework specific to the application should be designed using state-of-the-art DL principles to enhance the desired performance. Previous works, while showed feasibility, mostly utilized already-existing networks designed for other applications (e.g., denoising). Hence, elaborate image-to-image translation techniques (e.g., architecture and loss function design, training strategies) should be considered to establish a framework suited for the current application.

Here we present DL-based principles to accelerate the OCTA acquisition speed while addressing the above limitations. As mentioned before, our work differs from the previous studies in that we aim to tackle all factors related to the OCTA imaging speed (i.e., the B-scan repetition number and the sampling density), and thus enhance the pixel resolution and the image quality in a single integrated framework. By considering different model architectures, loss functions, and training strategies, we establish our networks using a two-staged adversarial training scheme to enhance the pixel resolution and the image quality successively. We evaluate our networks on an in-vivo mouse brain vasculature dataset and conduct a comparative study to select the best-performing model. Finally, we compare quantitative and qualitative results with baseline methods and examine our proposed framework’s reconstruction capability compared to the existing means.

Materials and methods

In this section, the OCTA dataset acquisition method is briefly described, followed by overall explanations of DL-based methodologies and the proposed approach’s workflow.

Data acquisition

Our dataset comprises in-vivo brain microvasculature images of 8-week-old male mice (C57BL/6J). After removing the scalp and skull (craniotomy details can be found in7), the samples were sealed with glass coverslips and imaged using a custom-designed spectral-domain OCT system from the University of Washington Seattle8. The system utilizes a broadband superluminescent diode as the light source with a central wavelength of 1340 nm and a 3 dB spectral bandwidth of 110 nm. A biaxial galvanometer is utilized to raster scan the probing light using a scanning step size of 10 μm, and a fast spectrometer detects the resulting interference signals at an A-line scan rate of 92 kHz. The system’s axial and lateral resolutions are 7 μm each, and a 3 mm × 3 mm field of view (FOV) was utilized. For each volumetric profile, 400 slow-axis locations were sampled at a B-scan rate of 180 Hz, containing cross-sectional structural B-scans with a pixel size of 400 × 1024 (400 A-lines per scan). As illustrated in Fig. 1, a previously reported OCTA algorithm7 was used to register eight sequential B-scans at each slow-axis location to generate a high-quality cross-sectional angiogram. We chose to utilize eight sequential B-scans since it is known that the angiogram quality (e.g., signal-to-noise ratio, vessel connectivity) saturates near eight repeated B-scans19. We downsampled the structural volumes in the enface plane, and two sequential downsampled B-scans were randomly selected to create a corresponding cross-sectional angiogram with low resolution and quality. The downsampling procedure was conducted to mimic the OCTA systems’ rasterized scanning principle. For example, if the artificial scanning step size is 2× larger than the fully-sampled step size, we downsample each axis in the enface plane by a ratio of two. The fully-sampled angiograms’ enface pixel resolution was 400 × 400, and we investigated downsampling ratios of 2×, 4×, and 8× in this study. For pre-processing, the pixel values were normalized in the range of [0, 1] to accelerate the deep neural network (DNN) training. In total, 29 pairs of volumetric angiograms were acquired from 11 different biosamples. The acquired volumetric angiograms were sectioned in 2D slices along the depth direction as enface representations. The enface angiograms from two sequential downsampled B-scans are low in resolution and quality. They are denoted as (LR, LQ) for the remainder of the paper and used as the input data to our DNN framework. Similarly, the high-resolution, high-quality enface angiograms from eight sequential fully-sampled B-scans are denoted (HR, HQ) and used as the target for training and ground truth for evaluation. (LR, LQ) and (HR, HQ) angiograms were paired according to the downsampling ratio.

Figure 1.

Figure 1

Schematic of our deep learning (DL) framework for accelerated optical coherence tomography angiography (OCTA). (LR, LQ) low-resolution and low-quality, (HR, LQ) high-resolution and low-quality, (HR, HQ) high-resolution and high quality, SSIM structural similarity index measure, MS-SSIM multiscale structural similarity index measure, PSNR peak signal-to-noise ratio.

The MIP enface angiograms are conventionally utilized to visualize the vascular profile in angiography applications. However, the number of volumetric angiograms available in this study is limited, leading to an insufficient amount of MIP data (29 total) to establish our DNNs. To address the limited data issue, we used 2000 depth-wise sectioned enface angiograms from seven volumetric pairs as our training and validation data which were randomly split at a ratio of 9:1. As illustrated in Fig. 2, sectioned enface angiograms differ from MIP enface angiograms in that void spaces are common in the sectioned enface angiograms. Thus, we manually selected images with sufficient vessel profiles to use as our training and validation data. The training dataset was used for optimizing the stochastic gradient descent algorithm, and the validation dataset was used during training to monitor the loss value and finetune the hyper-parameter settings. Random rotation and flipping in the x and y axes were applied to augment the training dataset. As for the test data, 200 sectioned enface angiograms were manually selected from an independent volumetric pair to conduct a comparative study and select the best-performing method. Furthermore, 21 MIP enface angiograms from the remaining volumetric pairs were utilized to evaluate the selected methods. A summary description of the dataset is provided in Table 1.

Figure 2.

Figure 2

Depiction of how depth-wise sectioned enface angiograms differ from maximum intensity projection (MIP) enface angiograms. The left column illustrates the low-resolution, low-quality (LR, LQ) angiograms from two repeated downsampled B-scans (downsampling ratio of four). The right column illustrates the corresponding high-resolution, high-quality (HR, HQ) angiograms from eight repeated fully-sampled B-scans.

Table 1.

The in-vivo mouse brain vasculature dataset used in this study. MIP, maximum intensity projection.

Training/Validation Test
Number of volumetric pairs 7 1 21
Image type Sectioned enface Sectioned enface MIP enface
Image quantity 1800/200 200 21

Deep learning architectures

We have developed an end-to-end CNN framework to reconstruct (HR, HQ) enface angiograms from their (LR, LQ) counterparts. As shown in Fig. 3, our reconstruction model is structured in two-stages and contains two constituent modules that operate in series, i.e., the super-resolution module followed by the quality-enhancing module. The super-resolution module compensates the pixel undersampling, and the quality-enhancing module compensates the B-scan repetition number. Our goal is to establish an end-to-end framework consisting of two sub-modules performing each task successively. The two-stage design should and has in our experiments been found to demonstrate better reconstruction performance than directly mapping (LR, LQ) angiograms to their (HR, HQ) counterparts. Further details on the proposed structure of the networks are summarized in Supplementary Table 1.

Figure 3.

Figure 3

Architectural description of the deep learning (DL) models used in this study: a dense connection-based model (a) and a residual connection-based model (b). (LR, LQ) low-resolution and low-quality, (HR, LQ) high-resolution and low-quality, (HR, HQ) high-resolution and high quality, PReLU parametric rectified linear unit.

The super-resolution module first upsamples the pixel resolution of the undersampled, low quality angiograms (LR, LQ) to their fully-sampled, low-quality counterparts (HR, LQ). Note that (LR, LQ) and (HR, LQ) angiograms are constructed from two repeated B-scans and, therefore, of poor quality. Thus, the task of mapping (LR, LQ) to (HR, LQ) angiograms is conducting pixel-wise super-resolution to fill in the missing pixels from undersampling. The module first extracts low-level features from the input angiograms using (9 × 9) convolution filters and then extracts high-level residuals using a series of repeated operations within the Deep layer. The Deep layer consists of five convolution blocks with (3 × 3) convolution filters and the parametric rectified linear unit (PReLU) activation function20. Convolution operations with a smaller receptive field are utilized to capture finer details of the input feature maps, and the PReLU function is employed as our primary activation function since it is known to accelerate deeper networks’ training by incorporating a trainable leakage coefficient20.

Two types of inter-block connections within the Deep layer are adopted: the dense connection21 (Fig. 3a) and the residual connection22 (Fig. 3b). We have adopted these inter-connection strategies as they have presented promising results for various previously reported bio-imaging applications17,18,23. The dense connection concatenates all subsequent convolution blocks’ output with the original input and utilizes the result as the input to the next convolution block. This configuration strengthens feature propagation since the convolution blocks only need to learn refinements that augment the previous outputs. The residual connection sums the input with the convolution block’s output, which is advantageous as the learning is alleviated to focus on the residuals. Also, these inter-block connections prevent the vanishing gradient problem and reduce the number of parameters for deeper networks. The extracted low-level and high-level features are merged using an element-wise summation operation. For the upsampling layer, we utilize the pixel shuffle operation24 to avoid the checkerboard artifacts that are commonly observed when using transposed convolutions. Unlike the previous approach, where the input image is pre-upsampled using filter-based methods18,23, we let the network learn the upsampling filters directly to increase performance25.

The quality enhancing module enhances the fully-sampled, low-quality output from the previous module and reconstructs the fully-sampled, high-quality angiograms (HR, HQ). The module specifically compensates for the B-scan repetition number (from two repeated B-scans to eight) and thus enhances the angiograms quality. It is worth mentioning that the feature map size is preserved throughout the module as the input and the ground truth (HR, HQ) angiograms are spatially aligned. For the most part, the input and the ground truth angiograms share similar low-frequency information but differ in the high-frequency details. Thus, we designed our module with (9 × 9) convolution operation to extract the low-frequency information. The operations within the Deep layer focus on reconstructing the high-frequency residuals. The two components are merged using the element-wise summation operation. Finally, a 400 × 400 reconstructed angiogram is obtained through a (9 × 9) convolution at the end of the module.

As mentioned before, the modules are concatenated in series to form a single reconstruction network. We establish two comparative network models based on the inter-block connection type adopted in the Deep layer: the DenseReconstNet (dense connection) and the ResReconstNet (residual connection).

Model training and evaluation

Several training strategies were employed to train our networks and optimize their trainable parameters. For the loss function L, we diverge from the de facto mean squared error (MSE) loss used in the previous studies17,18 and instead propose a perceptually improved loss function by combining different loss terms. While the MSE loss achieves high PSNR, minimizing the MSE loss encourages reconstructing pixel-wise averages of plausible outcomes, typically over-smooth and thus failing to capture the high-frequency details of small vessel profiles25. We utilize the mean absolute error (MAE) loss as the primary loss function since it is known to have better local convexity26:

LMAE=1Ni=1NY-Y^. 1

N denotes the number of pixels in each angiogram, Y and Y^ denotes the ground truth and the reconstructed angiogram, respectively. In addition to the primary loss function, we incorporate two other loss terms to aid in reconstructing the high-frequency details:

LMS-SSIM=1-MSSSIM(Y,Y^), 2
LFMAE=1Ni=1NF(Y)-F(Y^). 3

The first term is the multiscale structural similarity (MS-SSIM27) loss, with the MSSSIM(·) function calculating the corresponding metric. The MS-SSIM loss is advantageous in that it adeptly preserves the contrast in high-frequency regions when utilized along with the MAE loss26. The second term is the Fourier MAE loss (FMAE), which calculates the MAE from the magnitude of the angiograms’ 2D Fourier transforms. We choose FMAE to provide information on the vessel orientations and perceive the downsampling remnants as frequency corruptions23. The general strategy in incorporating different loss terms is that the primary loss function reconstructs the overall structure and the angiograms’ low-frequency components. In contrast, the two sub-loss terms reconstruct the high-frequency details and remove the angiograms’ frequency-space artifacts. In total, our loss function L is designed by linearly combining counterparts and is defined as follows:

L=LMAE+0.8×LMS-SSIM+7×10-5×LFMAE. 4

All the terms included in our loss function are fully differentiable, which is necessary for the neural network’s backpropagation algorithm. The weighting values for the different loss terms were referenced from previous literature23,26 and finetuned by heuristic searching and were found to be sufficient for all of the tested models. We mention that our proposed loss function should and has in our experiments been found to provide improved reconstruction performance and training stability than solely using the MAE loss. Each module is sequentially trained during each training iteration. The super-resolution module is trained first, followed by the quality enhancing module using the previous module’s output. At the end of each iteration, the two modules are concatenated as an end-to-end network and finetuned using a lower learning rate. Trainable parameters were initialized using the He normal initialization method20 and optimized using the Adam optimizer28. Early stopping and L2 regularization29 was also employed to avoid over-fitting the network parameters.

After successfully establishing the DenseReconstNet and the ResReconstNet, we further trained each network using an adversarial training scheme to boost the reconstruction performance. As illustrated in Fig. 4, we define a discriminator network D, which we optimized in an alternating manner with our pre-trained reconstruction model G to solve the adversarial min-max problem30:

minGmaxDEYpdata(Y)[logD(Y)]+EXpdata(X)[log(1-D(G(X)))]. 5

Figure 4.

Figure 4

Architectural description of the discriminator network used for the adversarial training. (HR, HQ), high-resolution and high-quality; LReLU, leaky rectified linear unit; and BN, batch normalization.

X denotes the (LR, LQ) angiogram taken as the input to our reconstruction model. The idea is that we train our reconstruction model to fool the discriminator that distinguishes the reconstructed angiograms from their (HR, HQ) counterparts. The proposed training strategy allows our reconstruction model to create perceptually superior solutions residing in the manifold of the (HR, HQ) angiograms. The adversarial loss function for the reconstruction models is defined as follows30:

LAdv=L+5×10-3×(-logD(G(X))). 6

The weight value for the adversarial loss term was referenced from previous literature25 and finetuned by heuristic searching and was found to be sufficient for all of the tested models. The established generative adversarial networks (GANs) are denoted as DenseReconstGAN and ResRecontGAN. Details on the discriminator network’s structure can be found in Supplementary Table 2. All hyper-parameters were searched for by using the random search strategy31 and their details can be found in Supplementary Table 3.

As for the evaluation metric, we consider the perceptual indices of SSIM32, MS-SSIM27, and the PSNR to assess the reconstruction performance. In addition, we assess the angiograms’ contrast by calculating the root-mean-square (RMS) contrast33 and the vessel connectivity by calculating the ratio of connected flow pixels to the total number of skeletonized pixels18,19.

Results and discussion

In this section, we present and discuss the experimental results of our OCTA reconstruction method. We again emphasize that our task differs from image super-resolution in the strict sense because our method compensates for both the pixel undersampling (image resolution) and the B-scan repetition number (image quality). However, no previous work has attempted to address both factors in a single integrated framework. Therefore, we compare our method with existing super-resolution methods, including interpolation-based image upsampling methods (i.e., nearest-neighbor, bicubic, and Lanczos interpolation) and a previously reported DL-based super-resolution network for retinal OCT angiograms called the high-resolution angiogram reconstruction network (HARNet)18. Implementation details for HARNet were carefully examined from the previous literature18, and its hyper-parameters were optimized using the random search strategy31 (Supplementary Table 3). We first quantitatively compare the reconstruction performance of the presented methods using the test dataset consisting of sectioned enface angiograms. Once the methods showing the most promising results are identified, we further assess quantitative and qualitative results on the test dataset consisting of MIP enface angiograms from various in-vivo samples.

Quantitative comparison using sectioned enface angiograms

To emphasize the advantage of our proposed training strategy (i.e., two-staged training with the perceptual loss function followed by adversarial training), we conduct a full ablation study in addition to the comparative study with the baseline methods, and quantitative results are summarized in Table 2. We mention that although HARNet was originally proposed as a super-resolution framework, it was trained to map (LR, LQ) images to (HR, HQ) images in our experiments and thus learns to compensate for both the pixel undersampling (image resolution) and the B-scan repetition number (image quality). Quality enhancement filters (e.g., median, bilateral, vesselness) were not used for the interpolation-based methods since sectioned enface angiograms do not contain full vascular profiles (see Fig. 2), and applying such filters deteriorated the results. For the same reason, angiogram quality measures (i.e., the contrast and vessel connectivity) were not assessed, and only the image reconstruction metrics were considered (i.e., SSIM, MS-SSIM, and PSNR). However, an additional quality enhancement filter was applied for the interpolation-based method when evaluating the MIP enface angiograms in the following section, and all metrics were assessed.

Table 2.

Performance metrics of different methods evaluated on the test dataset composed of sectioned enface angiograms.

Ratio Metric Baseline Proposed
Interpolation Deep learning Single-stage with MAE Two-stage with MAE Proposed training strategy
Nearest Bicubic Lanczos HARNet18 ResNet DenseNet ResNet DenseNet ResGAN DenseGAN
r = 2 SSIM 0.740 ± 0.031 0.756±0.033 0.751 ± 0.031 0.794±0.023 0.809 ± 0.023 0.824 ± 0.025 0.889 ± 0.031 0.913 ± 0.021 0.910 ± 0.032 0.921±0.021
MS-SSIM 0.872 ± 0.012 0.879±0.012 0.878 ± 0.010 0.912±0.008 0.923 ± 0.006 0.928 ± 0.009 0.951 ± 0.021 0.972 ± 0.012 0.972 ± 0.011 0.975±0.010
PSNR [dB] 25.75 ± 3.16 26.54±3.16 26.37 ± 3.14 27.19±2.95 27.35 ± 2.87 27.38 ± 2.83 26.78 ± 3.31 29.07 ± 2.98 28.04 ± 3.20 29.30±2.93
r = 4 SSIM 0.706 ± 0.042 0.720±0.040 0.711 ± 0.044 0.744±0.059 0.745 ± 0.060 0.772 ± 0.060 0.780 ± 0.032 0.810 ± 0.041 0.797 ± 0.041 0.862±0.041
MS-SSIM 0.819 ± 0.021 0.834±0.021 0.831 ± 0.024 0.852±0.040 0.854 ± 0.046 0.863 ± 0.038 0.900 ± 0.014 0.913 ± 0.020 0.910 ± 0.013 0.937±0.021
PSNR [dB] 23.11 ± 3.46 23.91±3.43 23.76 ± 3.42 24.05±3.33 24.10 ± 3.27 24.27 ± 3.30 25.87 ± 3.36 25.28 ± 3.26 25.85 ± 3.36 26.00±3.24
r = 8 SSIM 0.678 ± 0.064 0.681±0.051 0.665 ± 0.053 0.713±0.058 0.725 ± 0.059 0.739 ± 0.062 0.726 ± 0.054 0.726 ± 0.051 0.755 ± 0.062 0.755±0.041
MS-SSIM 0.743 ± 0.051 0.762±0.052 0.757 ± 0.051 0.769±0.079 0.775 ± 0.080 0.790 ± 0.081 0.833 ± 0.042 0.868 ± 0.031 0.855 ± 0.041 0.892±0.020
PSNR [dB] 21.12 ± 3.79 21.80±3.73 21.64 ± 3.74 22.08±3.17 22.86 ± 3.14 22.94 ± 3.20 22.88 ± 3.77 24.55 ± 3.17 22.93 ± 3.72 25.58±3.01

Our deep learning models are denoted as ResNet, ResGAN, DenseNet, and DenseGAN in short. MAE mean absolute error, SSIM structural similarity index measure, MS-SSIM multiscale structural similarity index measure, PSNR peak signal-to-noise ratio.

The models trained with our proposed training strategy (ResReconstGAN and DenseReconstGAN) outperform all baseline methods, including the interpolation-based methods and the previously reported HARNet. When solely using the MAE loss, the networks trained with the two-staged training scheme show improved performance over those trained in a single-stage manner that directly maps (LR, LQ) angiograms to their (HR, HQ) counterparts, including HARNet. The two-staged training strategy enhances the reconstruction performance since it provides additional intermediate supervision during training to perform super-resolution to (HR, LQ) angiograms. Furthermore, Table 2 shows that training with our proposed perceptual loss function and the adversarial training scheme further boosts the reconstruction performance as the metrics improved over the two-staged networks trained solely using the MAE loss. The results follow the intuition that the sub-loss terms and the adversarial loss push the reconstructed results to the (HR, HQ) angiograms’ manifold. The dense connection-based models (DenseReconstNet and DenseReconstGAN) outperform the residual connection-based models (HARNet, ResReconstNet, and ResReconstGAN), which is consistent with the consensus that deeper networks benefit from their ability to model mappings of higher complexity34,35. It is worth mentioning that the previously reported HARNet shows poor reconstruction performance even compared to our single-stage models trained using the MAE loss. One possible reason could be that while it is known that pre-upsampling the input image using filter-based methods deteriorates the results25, our networks learn the upsampling filters directly in contrast to HARNet. DenseReconstGAN consistently shows the best performance amongst all comparative methods for each of the downsampling ratios. HARNet shows the best performance amongst the baseline methods, and the bicubic interpolation method outperforms all other interpolation-based methods. Therefore, we focus our analysis on comparing the three methods for assessing the MIP test dataset, and the results are presented in the following section.

Quantitative and qualitative comparison using MIP enface angiograms

We compare the performance of the DenseReconstGAN, HARNet, and the bicubic interpolation method for the test dataset consisting of MIP enface images taken from various biosamples. The quantitative results of the three comparison methods are presented in Table 3. A bilateral filter was applied to the bicubic interpolation results to enhance the angiogram quality for fair comparisons with the DL-based methods. The results indicate that our DenseReconstGAN is superior to the baseline methods in terms of image reconstruction measures (i.e., SSIM, MS-SSIM, and PSNR) and angiogram quality measures (i.e., the contrast and vessel connectivity) for all downsampling ratios. Also, the results show that our DNN, trained with sectioned enface images, can generalize to full vascular profiles in MIP enface images. However, we notice that all evaluation metrics using the MIP enface dataset are lower than the sectioned enface dataset (see Table 2), especially for higher downsampling ratios of four and eight. The main reason is that the void regions in the sectioned enface angiograms yield high evaluation metrics, for they do not include contents for reconstruction (see Fig. 2). In addition, assessing the MIP enface angiograms require higher generalizability, as they are taken from various in-vivo samples. Nevertheless, our DenseReconstGAN still exhibits notable metrics at downsampling ratios of two and four and significantly outperforms the baseline methods, including the previously reported HARNet, which confirms the superiority of our proposed DL-based framework.

Table 3.

Performance metrics of our DenseReconstGAN, HARNet, and the bicubic interpolation method with bilateral filtering evaluated on the test dataset composed of maximum intensity projection (MIP) enface angiograms from various biosamples.

Ratio Metric Bicubic + Bilateral HARNet18 DenseReconstGAN
r = 2 SSIM 0.518 ± 0.022 0.779 ± 0.057 0.904±0.068
MS-SSIM 0.862 ± 0.020 0.900 ± 0.044 0.965±0.044
PSNR [dB] 15.17 ± 0.79 21.23 ± 1.47 25.40±2.31
Contrast 0.226 ± 0.007 0.266 ± 0.007 0.270±0.007
Connectivity 0.834 ± 0.014 0.868 ± 0.010 0.877±0.008
FPS 93.46 21.06 21.81
r = 4 SSIM 0.281 ± 0.029 0.617 ± 0.061 0.730±0.046
MS-SSIM 0.618 ± 0.038 0.802 ± 0.058 0.881±0.036
PSNR [dB] 12.56 ± 0.85 18.34 ± 1.47 20.21±1.44
Contrast 0.205 ± 0.010 0.254 ± 0.008 0.264±0.009
Connectivity 0.665 ± 0.040 0.864 ± 0.009 0.866±0.010
FPS 96.56 23.81 23.78
r = 8 SSIM 0.188 ± 0.021 0.347 ± 0.065 0.408±0.048
MS-SSIM 0.356 ± 0.032 0.582 ± 0.068 0.679±0.054
PSNR [dB] 10.92 ± 0.60 14.89 ± 1.19 15.86±1.15
Contrast 0.198 ± 0.006 0.216 ± 0.011 0.245±0.009
Connectivity 0.651 ± 0.029 0.727 ± 0.018 0.762±0.021
FPS 99.75 23.95 21.81

SSIM structural similarity index measure, MS-SSIM multiscale structural similarity index measure, PSNR peak signal-to-noise ratio, FPS frames per second.

Significance values are given in bold.

Figure 5 illustrates the qualitative performance of our DenseReconstGAN and HARNet on an (LR, LQ) MIP enface angiogram with a downsampling ratio of two. In the enlarged regions of interest (ROI), it can be seen that HARNet suffers from contrast loss and oversmooths the vascular details, resulting in dilated vessel structures. In contrast, our DenseReconstGAN demonstrates improved reconstruction precision with obvious advantages in recovering the high-frequency microvascular details owing to our improved training strategy. Since capillary morphology provides critical information about vascular pathology (e.g., capillary telangiectasias), we specifically compare the reconstruction performance of very small vessels ( 50 μm in diameter) along the dashed lines within each ROI. It can be observed that the input (LR, LQ) angiogram exhibits distorted and out-of-phase vessel profiles due to spatial aliasing. HARNet fails to distinguish vessels of smaller diameters (vessel profiles in the distance range of 100–125 μm, 150–175 μm for the red ROI, and 75–100 μm blue ROI) and severely loses contrast (0–25 μm red ROI and 150–200 μm blue ROI). Conversely, DenseReconstGAN reconstructs in-phase microvascular profiles with relatively small contrast loss. In addition, the vessel profiles indicate that our DenseReconstGAN produces no false blood flow signals. These results are better reflected in the angiograms’ contrast and vessel connectivity values in Table 3, as DenseReconstGAN shows superior values for all downsampling ratios. Some minor limitations can be observed using our method in certain poorly sampled regions. For example, local smoothing (115–150 μm red ROI) is observed in regions where spatial aliasing is most severe in the input (LR, LQ) angiogram. In addition, DenseReconstGAN also suffers from wavy motion artifacts, as can be observed in the red ROI. Our DL model is more prone to these artifacts since high expression capability, unfortunately, enhances noise. Our framework does not make efforts to correct these artifacts since commercial systems can easily address these artifacts by tracking the scan acquisition level and utilizing related software as described in Refs.36,37. Despite the above limitations, the qualitative results indicate that our proposed DL framework exhibits significantly superior performance than the bicubic interpolation method with bilateral filtering and the previously reported HARNet in reconstructing the microvascular profiles.

Figure 5.

Figure 5

Qualitative performance of our DenseReconstGAN and HARNet for reconstructing at a downsampling ratio of two. Enlarged profiles of the boxed regions of interest (ROI) are illustrated. Vessel profiles along the dashed lines within each ROI are also illustrated. (LR, LQ) low-resolution and low-quality, (HR, HQ) high-resolution and high quality.

We present a qualitative comparison of the two methods at higher downsampling ratios of four and eight (Fig. 6). A clear distinction can be made regarding the quality of vesselness, as the difference between the two methods becomes more significant with sparse data. At a downsampling ratio of four, oversmoothing and contrast loss worsens for HARNet, whereas microvascular structure and contrast are reasonably preserved for our DenseReconstGAN. Similarly, at a downsampling ratio of eight, HARNet exhibits biologically dubious features (jagged and disjointed vessels), while our DenseReconstGAN reconstructs smoother and rounder vessel profiles. However, at such a high downsampling ratio, the performance of our DL model deteriorates severely. Since the downsampling ratio of eight is equivalent to a scanning step size of 80 μm, and the scanning step size far exceeds the maximum Nyquist limit required to capture the capillary resolution ( 10 μm38), the degraded results are expected. Thus, the desired application context should always be considered when selecting the appropriate scanning step size to meet the desired performance. For example, a large scanning step size (e.g., 40 μm) is more appropriate for clinical applications requiring fast imaging speed but tolerant to the low image quality. Nevertheless, our DL framework still outperforms the previously reported super-resolution network HARNet in reconstructing the vasculature’s bulk physiology when using larger scanning step sizes of 40 and 80 μm. Thus, our method expands the range of usable scanning step sizes in OCTA systems, which were considered unavailable previously.

Figure 6.

Figure 6

Qualitative performance of our DenseReconstGAN and HARNet for reconstructing at downsampling ratios (denoted as r) of four and eight. (LR, LQ) low-resolution and low-quality, (HR, HQ) high-resolution and high quality.

We also compare the computation time and the mean frames per second (FPS) rate is provided in Table 3 (CPU and GPU details are listed in Supplementary Table 3). Even though our DL model is very deep with a high parameter dimension, it achieves a reasonably fast computation time during inference ( 23.78 FPS) because the forward path of a trained DNN model comprises only basic operations and nonlinear activation functions. HARNet exhibits a similar inference speed ( 23.95 FPS) since it is also a very deep model with a similar parameter dimension. Although the bicubic interpolation method with bilateral filtering offers the fastest computation time ( 99.75 FPS), it shows the worst reconstruction performance. Our DenseReconstGAN shows the best reconstruction results while offering computation speed sufficient for real-time embedded software usage. The computation time does not increase with the downsampling ratio, indicating that its consideration can be ignored when choosing the scanning step size. While (HR, HQ) OCTAs’ mean acquisition time was 17.8 s, the acquisition time reduced significantly for the (LR, LQ) OCTA case. For example, doubling the scanning step size (r = 2) and using two sequential B-scans would result in a 16× faster acquisition time, i.e., approximately 1.1 s. Similarly, ratios of four and eight would result in 64× and 256× faster acquisition time, i.e., 0.28 and 0.07 seconds, respectively. Thus, we believe our DL-based reconstruction framework is superior to the previously reported methods as it significantly reduces the (HR, HQ) OCTA’s acquisition time ( 1.2 s) while presenting enhanced reconstruction performance.

Conclusions

This study proposed a deep learning-based software-only solution to accelerate the OCTA systems’ imaging speed while preserving the image quality. The main contributions of the work are summarized as follows. First, we tackled all existing inherent factors related to the slow OCTA imaging speed, including the B-scan repetition number and the sampling density. While previous works only partially addressed the limiting factors, our work considered all aspects in a single, end-to-end framework. As a result, our approach achieved an acceleration of OCTA imaging by factors of up to 16–256×. Our method is advantageous since it can be directly applied to clinically available systems without increasing the system complexity. In addition, larger FOVs can be captured to improve clinical observations without sacrificing spatial resolution. Second, we established a novel DNN framework by leveraging state-of-the-art deep learning principles (i.e., network architectures, loss functions, and training strategies). In particular, we showed that the DenseReconstGAN exhibited the best performance and significantly outperformed the baseline methods, including the previously reported HARNet, in both quantitative and qualitative aspects for all of the experimented cases (Tables 2, 3, Figs. 5, 6). The results indicate that our proposed DNN framework is more suitable for the current application of compensating both the B-scan repetition number and the sampling density.

In the future, we aim to continually refine our DL framework’s generalizability by training with more MIP enface images from various in-vivo samples. We also aim to extend our framework to other pathological applications (e.g., retinal angiography). By combining our established framework with transfer learning techniques39, we can avoid acquiring a large amount of data that is required for re-training. We hope to utilize our integrated framework to broaden OCTA systems’ clinical applicability to rapidly and accurately delineate pathological biomarkers and achieve potential near real-time functional imaging capability.

Supplementary Information

Supplementary Tables. (39.1KB, docx)

Acknowledgements

The research of S.L and G.K was supported in part by the National Research Foundation of Korea (NRF) Grant funded by the Korea Government Ministry of Science and ICT (MSIT) (No. 2020R1A2C1009744), by Institute for Information communications Technology Promotion (IITP) Grant funded by the Korea government (MSIP) (No. 2019-0-01906, Artificial Intelligence Graduate School Program (POSTECH)), and the Priority Research Centers Program through the National Research Foundation of Korea (NRF) funded by the Ministry of Education (No. 2020R1A6A1A03047902). The research of J.K and C.K was supported in part by the National Research Foundation of Korea (NRF) Grant funded by the Korean government (MSIT) (No. NRF-2019R1A2C2006269) and Basic Science Research Program through the NRF funded by the Ministry of Education (No. 2020R1A6A1A03047902). It was also supported by the BK21 Four project. The research of W.J.C was supported in part by the Engineering Research Center (ERC) (No. 2020R1A5A1018052).

Author contributions

Conceptualization: C.K., J.K. Methodology: S.L., C.K., G.K., J.K. Investigation: G.K., J.K. Visualization: G.K., J.K. Supervision: S.L., C.K., W.J.C. Writing—original draft: G.K., J.K. Writing—review & editing: S.L., C.K., W.J.C., G.K., J.K.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

These authors contributed equally: Gyuwon Kim and Jongbeom Kim.

Contributor Information

Woo June Choi, Email: cecc78@cau.ac.kr.

Chulhong Kim, Email: chulhong@postech.edu.

Seungchul Lee, Email: seunglee@postech.ac.kr.

Supplementary Information

The online version contains supplementary material available at 10.1038/s41598-022-05281-0.

References

  • 1.Huang D, et al. Optical coherence tomography. Science. 1991;254:1178–1181. doi: 10.1126/science.1957169. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Drexler W, et al. In vivo ultrahigh-resolution optical coherence tomography. Opt. Lett. 1999;24:1221–1223. doi: 10.1364/OL.24.001221. [DOI] [PubMed] [Google Scholar]
  • 3.Bezerra HG, Costa MA, Guagliumi G, Rollins AM, Simon DI. Intracoronary optical coherence tomography: A comprehensive review: Clinical and research applications. JACC Cardiovasc. Interv. 2009;2:1035–1046. doi: 10.1016/j.jcin.2009.06.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Hammes H-P, Feng Y, Pfister F, Brownlee M. Diabetic retinopathy: Targeting vasoregression. Diabetes. 2011;60:9–16. doi: 10.2337/db10-0454. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Sattler EC, Kästle R, Welzel J. Optical coherence tomography in dermatology. J. Biomed. Opt. 2013;18:061224. doi: 10.1117/1.JBO.18.6.061224. [DOI] [PubMed] [Google Scholar]
  • 6.De Carlo TE, Romano A, Waheed NK, Duker JS. A review of optical coherence tomography angiography (OCTA) Int. J. Retin. Vitreous. 2015;1:1–15. doi: 10.1186/s40942-015-0005-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Choi WJ, Li Y, Wang R. Monitoring acute stroke progression: Multi-parametric OCT imaging of cortical perfusion, flow, and tissue scattering in a mouse model of permanent focal ischemia. IEEE Trans. Med. Imaging. 2019;38:1427–1437. doi: 10.1109/TMI.2019.2895779. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Choi WJ, Zhi Z, Wang R. In vivo OCT microangiography of rodent iris. Opt. Lett. 2014;39:2455–2458. doi: 10.1364/OL.39.002455. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Choi WJ. Optical coherence tomography angiography in preclinical neuroimaging. Biomed. Eng. Lett. 2019;9:311–325. doi: 10.1007/s13534-019-00118-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Fang PP, et al. Clinical applications of OCT angiography. Der Ophthalmol. Z. der Deutschen Ophthalmol. Gesellschaft. 2016;113:14–22. doi: 10.1007/s00347-015-0192-6. [DOI] [PubMed] [Google Scholar]
  • 11.Kim TS, et al. 9.4 MHz A-line rate optical coherence tomography at 1300 nm using a wavelength-swept laser based on stretched-pulse active mode-locking. Sci. Rep. 2020;10:1–9. doi: 10.1038/s41598-019-56847-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Klein T, et al. Multi-MHz retinal OCT. Biomed. Opt. Express. 2013;4:1890–1908. doi: 10.1364/BOE.4.001890. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Potsaid B, et al. Ultrahigh speed 1050 nm swept source/Fourier domain OCT retinal and anterior segment imaging at 100,000 to 400,000 axial scans per second. Opt. Express. 2010;18:20029–20048. doi: 10.1364/OE.18.020029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Fechtig DJ, Schmoll T, Grajciar B, Drexler W, Leitgeb RA. Line-field parallel swept source interferometric imaging at up to 1 MHz. Opt. Lett. 2014;39:5333–5336. doi: 10.1364/OL.39.005333. [DOI] [PubMed] [Google Scholar]
  • 15.Bonin T, Franke G, Hagen-Eggert M, Koch P, Hüttmann G. In vivo Fourier-domain full-field OCT of the human retina with 1.5 million A-lines/s. Opt. Lett. 2010;35:3432–3434. doi: 10.1364/OL.35.003432. [DOI] [PubMed] [Google Scholar]
  • 16.Potsaid B, et al. Ultrahigh speed spectral/Fourier domain OCT ophthalmic imaging at 70,000 to 312,500 axial scans per second. Opt. Express. 2008;16:15149–15169. doi: 10.1364/OE.16.015149. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Jiang Z, et al. Comparative study of deep learning models for optical coherence tomography angiography. Biomed. Opt. Express. 2020;11:1580–1597. doi: 10.1364/BOE.387807. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Gao M, et al. Reconstruction of high-resolution 6x6-mm OCT angiograms using deep learning. Biomed. Opt. Express. 2020;11:3585–3600. doi: 10.1364/BOE.394301. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Wang RK, et al. Wide-field optical coherence tomography angiography enabled by two repeated measurements of B-scans. Opt. Lett. 2016;41:2330–2333. doi: 10.1364/OL.41.002330. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.He, K., Zhang, X., Ren, S. & Sun, J. Delving deep into rectifiers: Surpassing human-level performance on imagenet classification. In Proc. IEEE International Conference on Computer Vision, 1026–1034 (2015).
  • 21.Huang, G., Liu, Z., Van Der Maaten, L. & Weinberger, K. Q. Densely connected convolutional networks. In Proc. IEEE Conference on Computer Vision and Pattern Recognition, 4700–4708 (2017).
  • 22.He, K., Zhang, X., Ren, S. & Sun, J. Deep residual learning for image recognition. In Proc. IEEE Conference on Computer Vision and Pattern Recognition, 770–778 (2016).
  • 23.DiSpirito A, III, et al. Reconstructing undersampled photoacoustic microscopy images using deep learning. IEEE Trans. Med. Imaging. 2021;40:562–570. doi: 10.1109/TMI.2020.3031541. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Shi, W. et al. Real-time single image and video super-resolution using an efficient sub-pixel convolutional neural network. In Proc. IEEE Conference on Computer Vision and Pattern Recognition, 1874–1883 (2016).
  • 25.Ledig, C. et al. Photo-realistic single image super-resolution using a generative adversarial network. In Proc. IEEE Conference on Computer Vision and Pattern Recognition, 4681–4690 (2017).
  • 26.Zhao H, Gallo O, Frosio I, Kautz J. Loss functions for image restoration with neural networks. IEEE Trans. Comput. Imaging. 2016;3:47–57. doi: 10.1109/TCI.2016.2644865. [DOI] [Google Scholar]
  • 27.Wang, Z., Simoncelli, E. P. & Bovik, A. C. Multiscale structural similarity for image quality assessment. In The Thrity-Seventh Asilomar Conference on Signals, Systems & Computers, 1398–1402 (2003).
  • 28.Kingma, D. P. & Ba, J. Adam: A method for stochastic optimization. Preprint at http://arxiv.org/abs/1412.6980 (2014).
  • 29.Goodfellow I, Bengio Y, Courville A, Bengio Y. Deep Learning. MIT Press; 2016. [Google Scholar]
  • 30.Goodfellow I, et al. Generative adversarial networks. Commun. ACM. 2020;63:139–144. doi: 10.1145/3422622. [DOI] [Google Scholar]
  • 31.Bergstra J, Bengio Y. Random search for hyper-parameter optimization. J. Mach. Learn. Res. 2012;13:281–305. [Google Scholar]
  • 32.Zhou W, Bovik AC, Sheikh HR, Simoncelli EP. Image quality assessment: From error visibility to structural similarity. IEEE Trans. Image Process. 2004;13:600–612. doi: 10.1109/TIP.2003.819861. [DOI] [PubMed] [Google Scholar]
  • 33.Peli E. Contrast in complex images. J. Opt. Soc. Am. A. 1990;7:2032–2040. doi: 10.1364/JOSAA.7.002032. [DOI] [PubMed] [Google Scholar]
  • 34.Szegedy, C. et al. Going deeper with convolutions. In Proc. IEEE Conference on Computer Vision and Pattern Recognition, 1–9 (2015).
  • 35.Simonyan, K. & Zisserman, A. Very deep convolutional networks for large-scale image recognition. Preprint at http://arxiv.org/abs/1409.1556 (2014).
  • 36.Camino A, Jia Y, Liu G, Wang J, Huang D. Regression-based algorithm for bulk motion subtraction in optical coherence tomography angiography. Biomed. Opt. Express. 2017;8:3053–3066. doi: 10.1364/BOE.8.003053. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Zang P, et al. Automated motion correction using parallel-strip registration for wide-field en face OCT angiogram. Biomed. Opt. Express. 2016;7:2823–2836. doi: 10.1364/BOE.7.002823. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Yao J, et al. High-speed label-free functional photoacoustic microscopy of mouse brain in action. Nat. Methods. 2015;12:407–410. doi: 10.1038/nmeth.3336. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Raghu, M., Zhang, C., Kleinberg, J. & Bengio, S. Transfusion: Understanding transfer learning for medical imaging. Preprint at http://arxiv.org/abs/1902.07208 (2019).