Table 5.

Least squares mean of differences between dialyzers over time of treatment, comparing the FX CorAL 600 versus SureFlux 17UX

LSMean, µg/L (95% CI) P Value
Parameter Pre 15 Min 60 Min 240 Min Pre 15 Min 60 Min 240 Min
C3a 6.66 (−21.07 to −34.40) 61.96 (34.23 to 89.70) 27.45 (−0.28 to 55.19) 21.42 (−6.31 to 49.16) 0.64 <0.001 0.05 0.13
C5a −0.011 (−0.039 to 0.017) 0.033 (0.005 to 0.061) −0.042 (−0.070 to −0.014) −0.005 (−0.033 to 0.023) 0.45 0.02 0.004 0.72
sC5b-9 1.07 (−18.09 to 20.23) 66.27 (47.12 to 85.43) 58.01 (38.86 to 77.17) 30.71 (11.56 to 49.87) 0.91 <0.001 <0.001 0.002

LSMean, least squares mean; 95% CI, 95% confidence interval of LSMean; pre, predialysis; 15 min, 15 minutes after start of HDF; 60 min, 60 minutes after start of HDF; 240 min, at the end of HDF; HDF, hemodiafiltration.