Table 3.
Sensitivity, specificity, false-positive (FP) value, false-negative (FN) value and area under the receiver operating characteristic curve (AUC) for evaluation of three-dimensional near-infrared reflection scans at 850 nm from the occlusal viewpoint (NIRR occlusal) and from trilateral evaluation (NIRR trilateral) and of digital bitewing radiography (BWR) using micro-computed tomography as a reference standard (lower and upper 0.95 confidence interval (CI) in parentheses)
Sensitivity | Specificity | FP | FN | AUC | ||
---|---|---|---|---|---|---|
NIRR occlusal | All samples | 0.33 (0.23–0.42) | 0.84 (0.78–0.89) | 0.16 (0.11–0.22) | 0.67 (0.57–0.74) | 0.58 (0.51–0.66) |
Premolars | 0.42 (0.26–0.58) | 0.89 (0.76–1.03) | 0.11 (− 0.03–0.24) | 0.58 (0.42–0.73) | 0.66 (0.51–0.80) | |
Molars | 0.26 (0.14–0.38) | 0.83 (0.76–0.89) | 0.17 (0.11–0.24) | 0.74 (0.62–0.81) | 0.54 (0.45–0.64) | |
NIRR trilateral | All samples | 0.47 (0.37–0.57) | 0.75 (0.69–0.82) | 0.25 (0.18–0.31) | 0.53 (0.43–0.61) | 0.61 (0.54–0.68) |
Premolars | 0.50 (0.34–0.66) | 0.79 (0.61–0.97) | 0.21 (0.03–0.39) | 0.50 (0.34–0.67) | 0.65 (0.50–0.79) | |
Molars | 0.44 (0.31–0.58) | 0.75 (0.68–0.82) | 0.25 (0.18–0.32) | 0.56 (0.42–0.64) | 0.60 (0.51–0.69) | |
BWR | All samples | 0.27 (0.17–0.36) | 1.00 (1.00–1.00) | 0.00 (0.00–0.00) | 0.73 (0.64–0.80) | 0.63 (0.55–0.70) |
Premolars | 0.33 (0.18–0.49) | 1.00 (1.00–1.00) | 0.00 (0.00–0.00) | 0.67 (0.51–0.81) | 0.65 (0.50–0.80) | |
Molars | 0.22 (0.11–0.33) | 1.00 (1.00–.100) | 0.00 (0.00–0.00) | 0.78 (0.67–0.84) | 0.60 (0.50–0.69) |