Table 2.
Accuracy of Corneal Astigmatism Measurements Stratified by Previous Refractive Surgery and Type of Astigmatism
Mean absolute difference |
Mean difference vector |
||||||
---|---|---|---|---|---|---|---|
OCT | Pentacam HR | IOLMaster 700 | OCT | Pentacam HR | IOL Master700 | ||
Net astigmatism | LASIK/PRK (n = 44) | 0.49 D | 0.69 D* | 0.61 D | 0.20 D @ 6° ± 0.55 D | 0.31 D @ 174° ± 0.75 D | 0.30 D @ 164° ± 0.72 D |
RK (n = 16) | 0.80 D | 1.18 D* | 1.28 D* | 0.23 D @ 3° ± 0.91 D | 0.32 D @ 87° ± 1.40 D | 0.68 D @ 105° ± 1.44 D | |
| |||||||
WTR (n = 17) | 0.40 D | 0.59 D | 0.66 D | 0.21 D @ 95° ± 0.46 D | 0.06 D @ 142° ± 0.72 D | 0.29 D @ 120° ± 0.90 D | |
ATR (n = 27) | 0.71 D | 0.90 D | 0.84 D | 0.47 D @ 4° ± 0.69 D | 0.14 D @ 175° ± 1.12 D | 0.22 D @ 119° ± 1.13 D | |
Plano (n = 13) | 0.47 D | 0.89 D | 0.91 D | 0.23 @ 176° ± 0.49 D | 0.39 D @ 173° ± 1.00 D | 0.42 D @ 162° ± 1.05 D | |
| |||||||
SimK astigmatism | LASIK/PRK (n = 44) | 0.79 D | 0.73 D | 0.67 D | 0.50 D @ 2° ± 0.73 D | 0.52 D @ 175° ± 0.67 D | 0.40 D @ 179° ± 0.68 D |
RK (n = 16) | 1.14 D | 1.03 D | 1.15 D | 0.15 D @ 164° ± 1.42 D | 0.33 D @ 123° ± 1.29 D | 0.44 D @ 96° ± 1.35 D | |
| |||||||
WTR (n = 17) | 0.60 D | 0.62 D | 0.60 D | 0.17 D @ 165° ± 0.64 D | 0.24 D @ 161° ± 0.70 D | 0.12 D @ 153° ± 0.75 D | |
ATR (n = 27) | 1.00 Da | 0.87 D | 0.80 D | 0.45 D @ 4° ± 1.14 D | 0.43 D @ 165° ± 1.01 D | 0.10 D @ 6° ± 1.04 D | |
None (n = 13) | 1.03 D | 0.85 D | 1.00 D | 0.64 D @ 179° ± 0.95 D | 0.59 D @ 173° ± 0.83 D | 0.55 D @ 173° ± 1.03 D |
The accuracy of corneal astigmatism measurements were measured by the vector difference with post-cataract surgery manifest refraction. LASIK = laser in situ keratomileusis; PRK = photorefractive keratectomy; RK = radial keratectomy; WTR = with-the-rule astigmatism; ATR = against-the-rule astigmatism; D = diopter.
P < .05 Comparison with OCT using generalized linear mixed-effect model.
P < .05 Comparison between eyes with WTR, ATR, or no refractive astigmatism using generalized linear mixed-effect model.