Table 2.

Performance of the CareStart Pf/Pan test compared to the gold standard, field microscopy

Field microscopy CareStart Pf/Pan
Positive Negative
Positive 96 0
Negative 4 100
Sensitivity, % (95% CI) 100 (96.23–100)
Specificity, % (95% CI) 96.15 (90.44–98.99)
PPV, % (95% CI) 96 (90.18–98.40)
NPV, % (95% CI) 100 (N/A)
Accuracy, % (95% CI) 98 (94.96–99.45)
Cohen’s kappa (95% CI) 0.96 (0.92–0.99)