Figure 6. Bad single-color control due to fluorescent signal in negative population.

Figure 6.

An example where the fluorescent signal in the negative population is low indicating that the single-color control has been run properly (a). There is fluorescent signal (indicated in the red box) in the negative population indicating that there may be non-specific binding of the fluorophore (b). The BV711 single-color control should be prepared again and rerun.