Skip to main content
. Author manuscript; available in PMC: 2022 Jan 30.
Published in final edited form as: Curr Protoc Cytom. 2020 Jun;93(1):e74. doi: 10.1002/cpcy.74

Table 4.

Troubleshooting problems and solutions

Problem Potential Cause Potential Solution
No cells on flow cytometer Cells were knocked out of the plate when performing the wash steps
Cytometer is clogged
When decanting supernatant, tap more gently
Run 10% bleach or 33% Contrad through the cytometer until the clog is removed
Fluorescent signal in the negative population of the single-color bead control Non-specific binding of antibody to negative beads Wash the beads with FACS Staining Buffer again to remove non-specific binding on the negative beads
Spectral signature of single-color control does not match expected signature for the fluorophore Contamination of control with another fluorescent antibody
Tandem dye degraded
Re-prepare control
Replace tandem dye with new vial
High proportion of cells are dead Cells left in DNAse/liberase for too long
Cells left in RBC lysing buffer for too long
Centrifuged too long and too hard
Cells incubated with Protein Transport Inhibitor for too long or too high of a concentration
Reduce the amount of time in the DNAse/liberase
Reduce the amount of time in the RBC lysing buffer
Reduce the centrifuge speed and/or time
Reduce the amount of time or concentration of Protein Transport Inhibitor
Fluorophore spill-over into other channels Spectra of two markers is indistinguishable Substitute a different fluorophore-marker pair
Low signal in certain markers Marker expression might be too low or inexistent
Experimental design (i.e., timepoint or vaccine stimulation) does not elicit certain markers
Markers chosen were too dim
Tandem dyes degraded or decoupled
Check that the marker is supposed to be expressed in the particular cells or animal model
Choose a time point closer to vaccination or stimulate the cells with PMA-ionomycin
Choose a brighter fluorophore for the specific marker or optimize dilution used
Use a new vial of the tandem dye
Not a clear separation in the live/dead sample stain Residual protein left in the media is binding to and quenching Zombie-NIR
Single-color control does not contain both positive and negative populations
Add additional washing step with PBS prior to staining with Zombie-NIR
Use ArC Amine Reactive Compensation Bead Kit (Thermo Fisher, cat# A10346) to run single-color live-dead control
Bad separation between negative and positive populations Antibody concentration is either too high or too low
Unbound antibodies were not adequately washed from the samples
Optimize dilution of antibodies
Add additional centrifugation and PBS wash