. 2022 Jan 31;27(1):016003. doi: 10.1117/1.JBO.27.1.016003

Table 5.

A comparison of the MPE predicted by ANSI Z136.1 for skin to our experimental results on blood vessel tissue. Note that the ANSI Z136.1 standard defines the MPE as 10× smaller than the dosage that will cause damage 50% of the time (EC50). Thus the MPE must be multiplied by 10 before comparing it to the applied dosage.

Condition Wavelength (nm) Experimentally applied fluence (J/cm2) Experimentally applied irradiance (W/cm2) 10× MPE fluence (J/cm2) 10× MPE irradiance (W/cm2) Expected damage from MPE (Y/N) Damage found experimentally (Y/N)
C1 1064 8.3 55 N N
C2 1064 7.8 10 N N
C3 1064 7.8 10 N Y
C4 1720 8.3 10 N N
C5 1720 2.6 1 Y Y
C6 1720 2.6 1 Y Y