Skip to main content
. 2022 Jan 6;163(3):bqac001. doi: 10.1210/endocr/bqac001

Table 3.

Accuracy in mouse muscle

Baseline Spiked Measured Accuracy
(pg/g) (pg/g) (pg/g) (%)
Estradiol
 Low 3.4 10.2 12.1 85
 High 3.4 81.5 86.6 102
Estrone
 Low 0.0 10.2 10.1 99
 High 0.0 81.5 79.2 97
Testosterone
 Low 26.6 127 166 109
 High 26.6 1019 1076 103
Dihydrotestosterone
 Low 0.0 25.5 28.1 110
 High 0.0 204 215 106
Progesterone
 Low 317 255 567 98
 High 317 2038 2298 97
Androstenedione
 Low 0.0 127 104 82
 High 0.0 509 479 94

Accuracy was calculated as (measured level – baseline level)/spiked level × 100%. Sample size: 125 mg.