Skip to main content
. 2022 Jan 6;163(3):bqac001. doi: 10.1210/endocr/bqac001

Table 5.

Accuracy in mouse gonadal fat

Baseline Spiked Measured Accuracy
(pg/g) (pg/g) (pg/g) (%)
Estradiol
 Low 0.0 7.2 7.8 108
 High 0.0 57.6 60.8 106
Estrone
 Low 0.0 7.2 7.9 109
 High 0.0 57.6 59.2 103
Testosterone
 Low 17.2 90.0 113 107
 High 17.2 720 756 103
Dihydrotestosterone
 Low 13.1 18.0 33.9 115
 High 13.1 144 165 106
Progesterone
 Low 3799 180 3968 94
 High 3799 1440 5226 99
Androstenedione
 Low 59.9 45.0 108 106
 High 59.9 360 427 102

Accuracy was calculated as (measured level – baseline level)/spiked level × 100%. Sample size 125 mg.