Table 2.

Pearson correlation coefficient, P values, 95% CI, and mean bias for heart rate (HR) and HR variability parameters between Ring and Shimmer3 in 5-minute window time.

Parameters Pearson correlation coefficient P value 95% CI Mean bias
HR 0.99341 <.001 −2.81 to 1.93 −0.44
RMSSDa 0.91502 <.001 −44.07 to 14.13 −14.97 ms
SDNNb 0.51772 <.001 −88.45 to 86.52 −0.96 ms
AVNNc 0.82486 <.001 −210.01 to 183.24 −13.39 ms
pNN50d 0.76024 <.001 −0.23 to 0.35 0.06
LFe band 0.42401 <.001 −1758.9 to 1806.12 23.61 ms2
HFf band 0.62734 <.001 −1423.92 to 1484.38 30.23 ms2
LF:HF ratio 0.35455 <.001 −2.53 to 2.31 −0.11

aRMSSD: root mean square of successive differences between normal heartbeats.

bSDNN: SD of normal beat-to-beat intervals.

cAVNN: average of normal heartbeat intervals.

dpNN50: percentage of successive beat-to-beat intervals that differ by more than 50 ms.

eLF: low frequency.

fHF: high frequency.