Table 1.
Category |
n (%) |
Intra-Op (%) |
Post-Op (%) |
---|---|---|---|
Total Adverse Events | 110 | 53 (48.2) | 57 (51.8) |
Total Device Issues | 63 (57.3) | 33 (52.4) | 30 (47.6) |
Break | 33 (52.4) | 27 (81.8) | 6 (18.2) |
Nonspecific break | 23 (69.7) | 19 (82.6) | 4 (17.4) |
Shoe Broke Off | 4 (12.1) | 4 (100) | 0 (0) |
Wire Broke from Piston | 1 (3) | 1 (100) | 0 (0) |
Shaft Broke from Wire | 1 (3) | 0 (0) | 1 (100) |
Shaft Broke Off | 2 (6.1) | 2 (100) | 0 (0) |
Phalange Broke Off | 1 (3) | 1 (100) | 0 (0) |
Disintegrated | 1 (3) | 0 (0) | 1 (100) |
Displacement | 17 (27) | 0 (0) | 17 (100) |
Nonspecific Displacement | 6 (35.3) | 0 (0) | 6 (100) |
Extrusion | 11 (64.7) | 0 (0) | 11 (100) |
Incorrect Size or Shape | 7 (11.1) | 3 (42.9) | 4 (57.1) |
Failure to Crimp | 4 (6.3) | 3 (75) | 1 (25) |
Loop of Stapes Open | 2 (3.2) | 0 (0) | 2 (100) |
Total Patient Issues | 39 (35.5) | 12 (30.8) | 27 (69.2) |
Hearing Loss | 12 (30.8) | 1 (8.3) | 11 (91.7) |
Distorted Hearing | 2 (5.1) | 0 (0) | 2 (100) |
Perforation | 3 (7.7) | 0 (0) | 3 (100) |
Erosion | 6 (15.4) | 0 (0) | 6 (100) |
Foreign Body in Patient | 4 (10.3) | 4 (100) | 0 (0) |
Nerve damage | 1 (2.6) | 1 (100) | 0 (0) |
Burns | 1 (2.6) | 1 (100) | 0 (0) |
Fracture | 1 (2.6) | 0 (0) | 1 (100) |
Pain | 2 (5.1) | 1 (50) | 1 (50) |
Inner Ear Injury | 1 (2.6) | 0 (0) | 1 (100) |
Vertigo | 1 (2.6) | 1 (100) | 0 (0) |
Dizziness | 1 (2.6) | 1 (100) | 0 (0) |
Headache and Pain During MRI | 1 (2.6) | 0 (0) | 1 (100) |
Drainage | 1 (2.6) | 0 (0) | 1 (100) |
Nausea | 1 (2.6) | 1 (100) | 0 (0) |
Nystagmus | 1 (2.6) | 1 (100) | 0 (0) |
Total Package Issues | 8 (7.3) | 8 (100) | 0 (0) |
Device Broken in Package | 2 (25) | 2 (100) | 0 (0) |
Contaminate Within Package | 2 (25) | 2 (100) | 0 (0) |
Unsealed Package | 1 (12.5) | 1 (100) | 0 (0) |
Improper Size Labeling | 1 (12.5) | 1 (100) | 0 (0) |
Incorrect Item Within Package | 1 (12.5) | 1 (100) | 0 (0) |
Missing Component | 1 (12.5) | 1 (100) | 0 (0) |