. 2021 Aug 28;17(1):18–24. doi: 10.1016/j.joto.2021.08.002

Table 6.

Subsequent intervention for device related events.

N/A Completed Procedure with Backup Device Prosthesis was Replaced Explanted Revision Tympanoplasty Piston Removed by Laser Surgery Unspecified Surgical Intervention Procedure not Completed
 Total Device Issues (n = 63) 12 (19) 22 (34.9) 18 (28.6) 6 (9.5) 2 (3.2) 1 (1.6) 1 (1.6) 1 (1.6)
 Break 6 (50) 17 (77.3) 5 (27.8) 3 (50) 0 (0) 1 (100) 0 (0) 1 (100)
 Nonspecific break 4 (66.7) 12 (70.6) 3 (60) 3 (100) 0 (0) 0 (0) 0 (0) 1 (100)
 Shoe Broke Off 1 (16.7) 3 (17.6) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Wire Broke from Piston 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 1 (100) 0 (0) 0 (0)
 Shaft Broke from Wire 0 (0) 0 (0) 1 (20) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Shaft Broke Off 0 (0) 2 (11.8) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Phalange Broke Off 1 (16.7) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Disintegrated 0 (0) 0 (0) 1 (20) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Displacement 4 (33.3) 0 (0) 9 (50) 1 (16.7) 2 (100) 0 (0) 1 (100) 0 (0)
 Nonspecific Displacement 0 (0) 0 (0) 5 (55.6) 1 (100) 0 (0) 0 (0) 0 (0) 0 (0)
 Extrusion 4 (100) 0 (0) 4 (44.4) 0 (0) 2 (100) 0 (0) 1 (100) 0 (0)
 Incorrect Size or Shape 1 (8.3) 3 (13.6) 3 (16.7) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Failure to Crimp 1 (8.3) 2 (9.1) 1 (5.6) 0 (0) 0 (0) 0 (0) 0 (0) 0 (0)
 Loop of Stapes Open 0 (0) 0 (0) 0 (0) 2 (33.3) 0 (0) 0 (0) 0 (0) 0 (0)