Table 8.

RCA for returned devices with device related events.

Cause Unknown Modification/Sizing/Mishandling Causing Break Met specifications Confirmed for Alleged Malfunction Product did not Meet Specifications In Progress
 Total Device Issues (n = 29) 6 (23.1) 7 (26.9) 6 (23.1) 4 (15.4) 2 (7.7) 1 (3.8)
 Break 4 (66.7) 7 (100) 1 (16.7) 4 (100) 2 (100) 0 (0)
 Nonspecific Break 4 (100) 4 (57.1) 0 (0) 0 (0) 2 (100) 0 (0)
 Shoe Broke Off 0 (0) 3 (42.9) 0 (0) 1 (25) 0 (0) 0 (0)
 Wire Broke from Piston 0 (0) 0 (0) 0 (0) 1 (25) 0 (0) 0 (0)
 Shaft Broke from Wire 0 (0) 0 (0) 1 (100) 0 (0) 0 (0) 0 (0)
 Shaft Broke Off 0 (0) 0 (0) 0 (0) 2 (50) 0 (0) 0 (0)
 Displacement 1 (16.7) 0 (0) 2 (33.3) 0 (0) 0 (0) 0 (0)
 Nonspecific Displacement 1 (100) 0 (0) 1 (50) 0 (0) 0 (0) 0 (0)
 Extrusion 0 (0) 0 (0) 1 (50) 0 (0) 0 (0) 0 (0)
 Incorrect Size or Shape 0 (0) 0 (0) 1 (16.7) 0 (0) 0 (0) 0 (0)
 Failure to Crimp 1 (16.7) 0 (0) 0 (0) 0 (0) 0 (0) 1 (100)
 Loop of Stapes Open 0 (0) 0 (0) 2 (33.3) 0 (0) 0 (0) 0 (0)