. 2001 Jun;39(6):2306–2307. doi: 10.1128/JCM.39.6.2306-2307.2001

TABLE 1.

Sensitivity and specificity of IDEIA PCE Chlamydia compared to LCx Chlamydia before and after confirmation with the blocking reagent and discordance analysis

IDEIA PCE result No. of samples with the following LCx result:
% Sensitivity % Specificity
Positive Negative
Before confirmation
 Positive 45 19 91.8 (45/49) 98.2 (1,055/1,074)
 Negative 4 1,055
After confirmation
 Positive 45 2b 91.8 (45/49) 99.8 (1,072/1,074)
 Negative 4a 1,072
a

The four false-negative IDEIA PCE samples were either positive for EBs or PCR positive. 

b

The two false-positive IDEIA PCE samples were negative for EBs and by PCR.