. 2022 Feb 4;24(4):294–308. doi: 10.1016/j.jmoldx.2021.12.010

Table 7.

An Overall Concordance Level of 91.89% Was Achieved When Comparing the SARS-CoV-2 Nucleocapsid LDT with the Wadsworth EUA

Wadsworth EUA results N1 nucleocapsid LDT N1 Wadsworth EUA N2 nucleocapsid LDT N2 Wadsworth EUA HRP nucleocapsid LDT HRP Wadsworth EUA Concordance between SARS-CoV-2 nucleocapsid LDT and Wadsworth EUA Notes
Early Ct (<30) 13.61 13.20 10.32 12.23 12.30 22.71 Concordant
 Positives 14.22 13.77 12.20 13.14 14.11 20.93 Concordant
14.51 14.03 13.08 13.31 15.08 21.67 Concordant
16.16 16.53 14.85 15.02 19.86 25.95 Concordant
16.40 16.86 15.23 16.59 20.37 25.32 Concordant
17.93 18.14 17.58 17.57 19.60 22.22 Concordant
18.62 19.56 16.75 17.38 22.78 24.15 Concordant
20.71 22.52 20.79 21.06 20.21 22.12 Concordant
23.09 24.32 21.03 23.77 24.71 25.33 Concordant
24.01 24.61 21.97 23.00 25.66 26.02 Concordant
23.47 24.56 21.50 23.38 23.44 24.47 Concordant
26.63 25.63 24.82 24.16 27.46 25.28 Concordant
24.62 27.28 22.80 25.64 26.79 20.19 Concordant
25.01 29.63 23.26 28.24 27.52 25.86 Concordant
27.21 29.87 27.25 31.19 23.89 24.84 Concordant
Late Ct (>30) 31.10 30.76 28.52 32.09 28.45 26.93 Concordant
 Positives 31.20 30.81 32.82 30.67 20.43 19.69 Concordant
27.37 31.35 27.60 30.01 27.02 26.07 Concordant
0.00 31.38 0.00 30.70 21.72 20.68 Discordant Negative in TaqPath EUA with KingFisher Flex Extraction
29.84 31.54 30.55 32.27 22.59 22.83 Concordant
29.45 31.82 29.22 31.14 28.22 27.37 Concordant
28.83 31.82 29.95 31.43 22.09 23.57 Concordant
32.17 32.03 31.41 30.25 20.42 22.13 Concordant
34.08 32.06 32.49 32.51 33.85 28.47 Concordant
30.43 32.44 31.42 34.39 22.82 22.30 Concordant
29.36 32.87 27.27 31.59 22.24 23.14 Concordant
34.50 32.89 34.94 35.17 24.30 26.15 Concordant
30.04 33.35 29.96 32.29 25.09 26.14 Concordant
0.00/0.00 34.00 0.00/0.00 35.89 22.24/21.62 21.77 Discordant X2 Negative in TaqPath EUA with KingFisher Flex Extraction
31.09 34.09 31.21 35.82 28.51 28.00 Concordant
33.87 34.58 0.00 33.62 22.92 24.19 Discordant Negative in TaqPath EUA with KingFisher Flex Extraction
32.95 35.05 36.34 35.12 25.81 27.29 Concordant
31.74 35.07 31.46 35.69 24.55 27.44 Concordant
33.84 35.11 31.65 34.78 28.27 30.11 Concordant
34.07 36.17 31.81 31.33 32.17 27.11 Concordant
32.87 35.52 36.40 34.41 22.23 24.22 Concordant
33.42 36.63 32.07 34.21 29.01 27.57 Concordant

A clinical evaluation was performed comparing the severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) nucleocapsid laboratory-developed test (LDT) with the Wadsworth Emergency Use Authorization (EUA). RNA was extracted by using the KingFisher Flex Extraction instrument with the MagMAX Viral/Pathogen II kit for the SARS-CoV-2 nucleocapsid LDT. RNA extraction for the Wadsworth EUA was performed with the EZ1 Advanced XL instrument and the EZ1 Virus Mini Kit 2.0. CT values and concordance results are listed for each patient sample. Notes for discordant samples are provided in the last column.

HRP, human RNAse P.