Skip to main content
. 2001 Jul;39(7):2572–2575. doi: 10.1128/JCM.39.7.2572-2575.2001

TABLE 1.

Sensitivities and specificities of tests for detection of antibodies to HIV-1 and HIV-2 with samples confirmed as positive or negative by WB assay

Test Sensitivity
Specificity
Serum samples (n = 65)
Plasma samples (n = 60)
Total sensitivity (%) (n = 125) Serum samples (n = 40)
Plasma samples (n = 35)
Total specificity (%) (n = 75)
No. negativea Sensitivity (%) No. negative Sensitivity (%) No. positiveb Specificity (%) No. positive Specificity (%)
Determine HIV-1/2 0 100 0 100 100 0 100 0 100 100
HIV SPOT 0 100 2 96.8 98.4 0 100 0 100 100
Innotest HIV-1/2 0 100 1 98.4 99.2 1 97.6 1 97.2 97.4
a

Number of samples (serum and plasma) detected as false negatives. 

b

Number of samples (serum and plasma) detected as false positives.