Failure to detect signal (Table 1, steps 1–2) |
a. Adjust camera focus deeper |
a. Focusing the camera at inappropriate level (e.g. at the dural surface) reduces signal amplitude (see “Methods”, “Imaging procedures” sections) |
b. Adjust camera position |
b. Cranial window must be positioned at the center of the camera, reflecting the brightest light |
c. Decrease/increase light intensity |
c. Hemodynamic signals evoke small reflectance intensity changes. Too little light may prevent signal detection |
d. Adjust anesthesia level |
d. High levels of anesthesia impair visual signals. It is important to pair sedation with minimal maintenance anesthesia during imaging |
Software or Image acquisition freezes/crashes |
a. Clear all or Restart MATLAB |
a. Clear workspace or restart MATLAB to flush memory |
b. Unplug/plug data acquisition (DAQ) device |
b. DAQ device may power off if software freezes, preventing additional recordings |
c. Reboot primary computer |
c. Extreme case, only when the experimenter is unable to acquire images |