. 2022 Jan 14;4(1):e210063. doi: 10.1148/rycan.210063

Figure 3:

Agreement between lesion diffusion-weighted imaging (DWI) measures by using multiband (MB) sensitivity encoding (SENSE) and conventional DWI (cDWI) techniques. Shown are Bland-Altman plots for (A) apparent diffusion coefficient (ADC) measures, (B) contrast-to-noise ratio (CNR) measures, and (C) signal index (SI) values. Bland-Altman analysis demonstrates no apparent systematic bias in the ADC, CNR, or SI measurements between the techniques. The dashed line indicates the mean ratio (MB SENSE DWI vs cDWI), and the dotted lines indicate the 95% limits of agreement.

Agreement between lesion diffusion-weighted imaging (DWI) measures by using multiband (MB) sensitivity encoding (SENSE) and conventional DWI (cDWI) techniques. Shown are Bland-Altman plots for (A) apparent diffusion coefficient (ADC) measures, (B) contrast-to-noise ratio (CNR) measures, and (C) signal index (SI) values. Bland-Altman analysis demonstrates no apparent systematic bias in the ADC, CNR, or SI measurements between the techniques. The dashed line indicates the mean ratio (MB SENSE DWI vs cDWI), and the dotted lines indicate the 95% limits of agreement.