TABLE 3.

As overview of the imaging performance of the IRm, the weighted cone‐beam dose index (CBDIw), contrast‐noise‐ratio (CNR) (for the LDPE, polystyrene, and Delrin® insert), non‐uniformities (determined from the lateral/anterior–posterior (AP) CTP486 computed tomography (CT)‐number profiles), image noise (from integrating the noise power spectrum (NPS)) converted to a CBDIw of 1 mGy, as well as the imaging fidelity, that were altogether measured for each investigated scan protocol, are listed

CNR Non‐uniformity (HU)
Protocol CBDIw (mGy) LDPE Polystyrene Delrin® Lateral AP Noise (HU) at 1 mGy Imaging fidelity (mm)
Pelvis
‐ Low dose (PL) 9.7 ± 0.3 10.4 ± 0.2 8.4 ± 0.3 14.8 ± 0.5 ‐6 ± 11 ‐6 ± 14 43.6 ± 0.6 0.18 ± 0.11
‐ Standard (PS) 13.6 ± 0.2 15.4 ± 0.2 10.2 ± 0.3 21.2 ± 0.8 2 ± 11 ‐34 ± 10 39.9 ± 0.7 0.21 ± 0.18
‐ Obese (PO) 16.8 ± 0.2 17.0 ± 0.4 10.3 ± 0.3 20.8 ± 0.7 ‐20 ± 10 ‐35 ± 7 41.3 ± 0.8 0.19 ± 0.16
‐ High‐quality (PH) 16.8 ± 0.2 21.5 ± 1.6 14.7 ± 1.7 28.1 ± 2.9 ‐2 ± 15 ‐3 ± 23 35.2 ± 1.3 0.39 ± 0.53
Breast
‐ Standard (BS) 5.0 ± 0.2 7.5 ± 0.5 5.6 ± 0.2 10.5 ± 0.5 ‐2 ± 12 ‐2 ± 17 43.8 ± 0.9 0.19 ± 0.23
‐ Obese (BO) 7.8 ± 0.3 10.4 ± 0.3 7.2 ± 0.3 14.1 ± 0.5 9 ± 10 12 ± 11 39.7 ± 0.9 0.17 ± 0.12
‐ High‐quality (BH) 8.7 ± 0.2 10.9 ± 0.2 8.3 ± 0.2 14.8 ± 0.4 1 ± 10 3 ± 18 39.7 ± 1.4 0.20 ± 0.15
Abdomen
‐ Low dose (AL) 2.0 ± 0.2 1.79 ± 0.04 1.41 ± 0.04 2.39 ± 0.05 120 ± 60 136 ± 19 115.8 ± 0.9 0.18 ± 0.14
‐ Standard (AS) 11.9 ± 0.4 4.23 ± 0.09 3.22 ± 0.06 5.67 ± 0.10 144 ± 16 131 ± 6 113.6 ± 0.2 0.18 ± 0.14
‐ Obese (AO) 40.6 ± 0.5 8.59 ± 0.09 6.25 ± 0.06 11.44 ± 0.15 ‐45 ± 10 ‐40 ± 5 122.9 ± 1.3 0.19 ± 0.18

Abbreviations: AP, anterior–posterior; HU, Hounsfield Unit.