Table 2.

Conventional MR brain tumor protocol used for 3 T.

Examination
Parameters
2D ax T2 FLAIR 2D T2 ax DWI ax 3D SWI ax 3D T1 ax pre 2D T2 cor PWI ax 3D T1 ax Post 3D FLAIR
TSE + IR TSE EPI-SE GRE MPRAGE TSE SS-EPI MPRAGE TSE + IR
Voxel dimensions 0.9 × 0.9 0.8 × 0.6 1.8 × 1.8 0.9 × 0.9 1 × 1 0.4 × 0.4 1.8 × 1.8 1 × 1 1 × 1
Matrix size 256 × 256 250 × 384 128 × 128 256 × 192 256 × 256 531 × 640 128 × 128 256 × 256 256 × 256
No. slices 36 40 30 80 192 56 19 192 176
Field of view [mm2] 230 210 230 230 220 230 230 220 250
Slice thickness, mm 4 3 5 1.75 1 3 5 1 0.9
TE [ms] 100 88 78 20 3.79 115 32 3.79 393
TI [ms] 2500 - - - 1100 - - - 2050
TR [ms] 9220 3490 4000 28 1800 4290 1400 1800 7000
TA, [min:s] 4:38 1:25 1:38 3:52 5:44 3:40 1:17 5:44 3:39
GRAPPA factor - - 2 2 - 2 2 - -
BW/pixel, [Hz/pixel] 170 199 1502 120 200 176 1346 200 651
FA [°] 150 120 - 15 12 120 90 12 T2var
Fat saturation yes No Yes No No No Yes No Yes

2D T2 FLAIR TSE, 2D T2 weighted fluid-attenuated inversion recovery turbo spin echo; 2D T2 TSE, 2D T2 weighted turbo spin echo; DWI, diffusion weighted echo-planar spin echo imaging; 3D SWI, 3D susceptibility weighted gradient recalled echo imaging; 3D T1 MPRAGE (pre-/post), 3D magnetization prepared rapid acquisition gradient echo sequence pre- and post-contrast injection, post-contrast images are collected with equivalent parameters to pre-contrast images; PWI, perfusion weighted imaging with single-shot echo planar imaging.