Table 3.

MRF protocols used for 3T.

Examination
Parameters
2D ax T2 FLAIR 2D T2 ax 3D T1 Sag 2D Multi-Echo
Spin Echo
MRF
TSE + IR TSE MP2RAGE
(T1 map)
(T2 map)
Voxel dimensions [mm2] 0.6 × 0.6 0.7 × 0.7 1.0 × 1.0 0.7 × 0.7 1.0 × 1.0
Matrix size 384 × 276 320 × 240 256 × 216 320 × 257 256 × 256
No. slices 10 23 160 10 10–13
Field of view [mm2] 230 × 166 230 × 170 256 × 216 230 × 180 256 × 256
Slice thickness [mm] 5.0 5.0 1.0 5.0 5.0
TE [ms] 126 118 2.98 12.6, 25.2, … 201.6 2.0
TI [ms] 2500 700, 2500 21.0
TR [ms] 8500 4890 5000 2100 12.14–15.00 (varied by sequence)
TA [min:sec] 3:43 1:25 8:02 3:38 3:51–4:51
Acceleration factor 1 (turbo factor: 19) 2 2 3 24 (inner k-space), 48 (outer k-space)
BW/pixel [Hz/pixel] 140 130 240 150 RX-Bandwidth: 400 kHz
FA [°] 180 180 4, 5 180 0–74 (varied by sequence)
Fat saturation Yes No No no

FLAIR fluid-attenuated inversion recovery; TSE, turbo spin echo; MP2RAGE, magnetization prepared 2 rapid acquisition gradient echoes; MRF, Magnetic Resonance Fingerprinting.