Skip to main content
. 2022 Jan 27;12:789731. doi: 10.3389/fmicb.2021.789731

TABLE 4.

Agreement rates regarding the resistance mechanism obtained in the intercenter evaluation for all centers participating in the study.

Resistance mechanism Diagnosis N %
Susceptible Agreement rate 51 94.40%
Minor errors 3 5.60%
Very major errors 0 0.00%
CIT Agreement rate 8 88.90%
Minor errors 1 11.10%
Very major errors 0 0.00%
SHV Agreement rate 8 88.90%
Minor errors 1 11.10%
Very major errors 0 0.00%
CMY-2 Agreement rate 17 94.40%
Minor errors 1 5.60%
Very major errors 0 0.00%
FOX Agreement rate 24 88.90%
Minor errors 3 11.10%
Very major errors 0 0.00%
K-1 Agreement rate 9 100.00%
Minor errors 0 0.00%
Very major errors 0 0.00%
CTX-M Agreement rate 120 95.20%
Minor errors 6 4.80%
Very major errors 0 0.00%
KPC Agreement rate 89 98.90%
Minor errors 0 0.00%
Very major errors 1 1.10%
IMP Agreement rate 26 96.30%
Minor errors 1 3.70%
Very major errors 0 0.00%
NDM Agreement rate 43 95.60%
Minor errors 0 0.00%
Very major errors 2 4.40%
VIM Agreement rate 79 87.80%
Minor errors 4 4.40%
Very major errors 7 7.80%
OXA-48-like Agreement rate 141 97.90%
Minor errors 3 2.10%
Very major errors 0 0.00%