Table 1.

Validation results of the recommended HPLC method for analysis of ALB, LEV ABA, and CLO.

Parameters ALB LEV ABA CLO
Range in (µg/mL) 15.15–93.75 25–150 30–150 11.7–140.63
Slope (mAU*s mL/µg) 32.252 23.239 4.174 18.312
Intercept (mAU*s) −9.008 −26.128 2.028 4.568
R2 0.99984 0.99980 0.99997 0.99990
Accuracy
(mean ± SD)
101.06 ± 0.489 101.15 ± 0.579 99.61 ± 0.588 100.58 ± 0.723
Precision (%RSD)
Repeatability * RSD ≤ 2%
Day to day precision **
0.47 0.33 0.35 1.25
Pooled RSD ≤ 3% 1.47 1.62 0.29 2.03
LOD in (µg/mL) *** 1.669 2.989 1.167 1.991
LOQ in (µg/mL) *** 5.056 9.057 3.537 6.033

* The intra-day precision (n = 6), average of one concentration repeated six times in the same day. ** The inter-day precision (n = 3), 3 replicates of a single sample were implemented on the first day, and then, on a second day, 3 replicates of freshly prepared samples were analyzed. The same analyst performed both tests. *** Limits for detection and quantitation were calculated via LOD = (SD of the response/slope) × 3; LOQ = (SD of the response/slope) × 10.