Table 1.
References | Study population | Results |
---|---|---|
Dunn (26) | 50 UV (at necropsy) BW 680–4,027 g, EG 26–44 weeks |
Shoulder-umbilicus distance |
Shukla and Ferrara (27) | 39 UV and 4 UA, BW 2,037 ± 1,077 g 10 UV, BW 2,260 ± 1,144 g |
Lenght (cm) = [(3 * BW in Kg + 9) / 2 +1] |
Verheij et al. (28) | 143 UV using the Shukla formula 125 UV using the revised formula |
Lenght (cm) = [(3 * BW in Kg + 9) / 2] |
Vali et al. (29) | 82 UV and 55 UA BW 1,311± 888 g, EG 23–27+6 weeks |
Umbilicus-mid-xiphoid-bed distance |
Gupta et al. (30) | 170 UV, BW 490 ± 4,800 g, EG 24–41 weeks 125 UA, BW 490 ± 4,800 g, EG 24–41 weeks |
Lenght (cm) = Umbilical-Nipple length – 1 cm |
The Shukla formula had the highest rate of either correct or high position. The Dunn formula performed quite poorly.