TABLE 1.
Allplex SARS‐CoV‐2/FluA/FluB/RSV kit | Standard M nCoV real‐time detection kit | U‐TOP COVID‐19 detection kit | |
---|---|---|---|
Target genes | S, RdRp, and N genes | E and RdRp genes | ORF1ab, N, S, and E genes |
Positive results | |||
Any target genes | 90 | 89 | 88 |
All target genes | 85 | 87 | 84 |
S and RdRP genes | 1 | ||
S gene | 3 | ||
RdRP gene | 1 | 2 | |
ORF1ab, N, and E genes | 1 | ||
ORF1ab and N genes | 1 | ||
ORF1ab and E genes | 1 | ||
E gene | 1 | ||
Negative results | 159 | 160 | 161 |
The positive results were defined the result of at least two of the three SARS‐CoV‐2 real‐time PCR assays.