. 2022 Jan 12;36(2):e24242. doi: 10.1002/jcla.24242

TABLE 3.

Details of discordant clinical sample results

Sample No. Reference standard Allplex SARS‐CoV‐2/FluA/FluB/RSV kit Standard M nCoV real‐time detection kit U‐TOP COVID‐19 detection kit
Result S (C t ) RdRP (C t ) N (C t ) Result RdRP (C t ) E (C t ) Result ORF1ab (C t ) N (C t ) S (C t ) E (C t )
1 Positive Target detected ND 37.3 ND Target detected 35.6 ND Target not detected ≥38.0 ≥38.0 ND ND
2 Positive Target detected 36.0 37.5 35.7 Target detected 35.7 34.1 Target detected 37.0 ≥38.0 ≥38.0 37.7
3 Positive Target detected 35.8 36.6 ND Target detected 34.0 34.4 Target detected 36.8 37.7 ≥38.0 ≥38.0
4 Positive Target detected 35.5 36.7 34.5 Target detected 35.6 34.0 Target detected 35.9 36.6 ND 35.4
5 Positive Target detected 36.5 ND ND Target detected 35.1 33.8 Target detected 36.0 37.0 37.6 36.1
6 Negative Target detected 37.4 ND ND Target not detected ND ND Target not detected ≥38.0 ≥38.0 ND ≥38.0

Abbreviations: C t , cycle threshold; ND, not detected.