TABLE 4.
SARS‐CoV‐2 PCR assays | Results according to reference standards | Sensitivity (95% CI) | Specificity (95% CI) | |||
---|---|---|---|---|---|---|
TP | FP | FN | TN | |||
Allplex SARS‐CoV‐2/FluA/FluB/RSV kit | 89 | 1 | 0 | 159 | 100 (95.94–100) | 99.38 (96.57–99.98) |
Standard M nCoV real‐time detection kit | 89 | 0 | 0 | 160 | 100 (95.94–100) | 100 (97.72–100) |
U‐TOP COVID‐19 detection kit | 88 | 0 | 1 | 160 | 98.88 (93.90–99.97) | 100 (97.72–100) |
Abbreviations: CI, confidence interval; FN, false negative; FP, false positive; TN, true negative; TP, true positive.