TABLE 4.

Clinical performance comparison of the three rRT‐PCRs for detection of SARS‐CoV‐2

SARS‐CoV‐2 PCR assays Results according to reference standards Sensitivity (95% CI) Specificity (95% CI)
TP FP FN TN
Allplex SARS‐CoV‐2/FluA/FluB/RSV kit 89 1 0 159 100 (95.94–100) 99.38 (96.57–99.98)
Standard M nCoV real‐time detection kit 89 0 0 160 100 (95.94–100) 100 (97.72–100)
U‐TOP COVID‐19 detection kit 88 0 1 160 98.88 (93.90–99.97) 100 (97.72–100)

Abbreviations: CI, confidence interval; FN, false negative; FP, false positive; TN, true negative; TP, true positive.